Manual
Table Of Contents
- 1 Safety Instructions
- 2 Applications
- 3 Documentation
- 4 Getting Started
- 5 The Instrument
- 6 Operating and Display Elements
- 7 Operation
- 8 Instrument Settings
- 9 Database
- 10 General Information on Measurements
- 10.1 Using Cable Sets and Test Probes
- 10.2 Test Plug – Changing Inserts
- 10.3 Connecting the Instrument
- 10.4 Automatic Settings, Monitoring and Shutdown
- 10.5 Measured Value Display and Memory
- 10.6 Help Function
- 10.7 Setting Parameters or Limit Values using RCD Measurement as an Example
- 10.8 Freely Selectable Parameter Settings or Limit Values
- 10.9 2-Pole Measurement with Rapid or Semiautomatic Polarity Reversal
- 11 Measuring Voltage and Frequency
- 12 Testing RCDs
- 12.1 Measuring Touch Voltage (with reference to nominal residual current) with ⅓ Nominal Residual Current and Tripping Test with Nominal Residual Current
- 12.2 Special Tests for Systems and RCDs
- 12.2.1 Testing Systems and RCCBs with Rising Residual Current (AC) for Type AC, A/F, B/B+ and EV/MI RCDs (PROFITEST MTECH+, PROFITEST MXTRA only)
- 12.2.2 Testing Systems and RCCBs with Rising Residual Current (AC) for Type B/B+ and EV/MI RCDs (PROFITEST MTECH+PROFITEST MXTRA)
- 12.2.3 Testing RCCBS with 5 × IDN
- 12.2.4 Testing of RCCBs which are Suitable for Pulsating DC Residual Current
- 12.3 Testing of Special RCDs
- 12.4 Testing Residual Current Circuit Breakers in TN-S Systems
- 12.5 Testing of RCD Protection in IT Systems with High Cable Capacitance (e.g. In Norway)
- 12.6 Testing of 6 mA Residual Current Devices RDC-DD/RCMB (RDC-DD: PROFITEST MXTRA and PROFITEST MTECH+ only)
- 13 Testing of Breaking Requirements for Overcurrent Protective Devices, Measurement of Loop Impedance and Determination of Short-Circuit Current (ZL-PE and ISC Functions)
- 14 Measuring Supply Impedance (ZL-N Function)
- 15 Earthing Resistance Measurement (Function RE)
- 15.1 Earthing Resistance Measurement – Mains Powered
- 15.2 Earthing Resistance Measurement – Battery Powered, “Battery Mode” (PROFITEST MPRO & PROFITEST MXTRA only)
- 15.3 Earthing Resistance, Mains Powered – 2-Pole Measurement with 2-Pole Adapter or Country-Specific Plug (Schuko) without Probe
- 15.4 Earthing Resistance Measurement. Mains Powered – 3-Pole Measurement: 2-Pole Adapter with Probe
- 15.5 Earthing Resistance Measurement, Mains Powered – Measuring Earth Electrode Potential (UE Function)
- 15.6 Earthing Resistance Measurement, Mains Powered – Selective Earthing Resistance Measurement with Current Clamp Sensor as Accessory
- 15.7 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – 3-Pole (PROFITEST MPRO & PROFITEST MXTRA only)
- 15.8 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – 4-Pole (PROFITEST MPRO & PROFITEST MXTRA only)
- 15.9 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Selective (4-pole) with Current Clamp Sensor and PRO-RE Measuring Adapter as Accessory (PROFITEST MPRO & PROFITEST MXTRA only)
- 15.10 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Ground Loop Measurement (with current clamp sensor and transformer, and pro-re measuring adapter as accessory) (PROFITEST MPRO & PROFITEST MXTRA only)
- 15.11 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Measurement of Soil Resistivity rE (PROFITEST MPRO & PROFITEST MXTRA only)
- 16 Measurement of Insulation Resistance
- 17 Measuring Low-Value Resistance of up to 200 W (Protective Conductor and Equipotential Bonding Conductor)
- 18 Measurement with Accessory Sensors
- 19 Special Functions – EXTRA Switch Position
- 19.1 Voltage Drop Measurement (at ZLN) – DU Function
- 19.2 Measuring the Impedance of Insulating Floors and Walls (standing surface insulation impedance) – ZST Function
- 19.3 Testing Meter Startup with Earthing Contact Plug – kWh Function
- 19.4 Leakage Current Measurement with PRO-AB Leakage Current Adapter as Accessory – IL Function (PROFITEST MXTRA only)
- 19.5 Testing Insulation Monitoring Devices – IMD Function (PROFITEST MXTRA only)
- 19.6 Residual Voltage Test – Ures Function (PROFITEST MXTRA only)
- 19.7 Intelligent Ramp – ta+ID Function (PROFITEST MXTRA only)
- 19.8 Testing Residual Current Monitors – RCM Function ( PROFITEST MXTRA only)
- 19.9 Checking the Operating Statuses of Electric Vehicles at Charging Stations per IEC 61851 ((PROFITEST MTECH+ & PROFITEST MXTRA)
- 19.10 PRCD – Test Sequences for Documenting Fault Simulations at PRCDs with the PROFITEST PRCD Adapter (PROFITEST MXTRA only)
- 20 Test Sequences (Automatic Test Sequences) – AUTO Function
- 21 Maintenance
- 22 Contact, Support and Service
- 23 CE Declaration
- 24 Disposal and Environmental Protection
- 25 Appendix
- 25.1 Tables for Determining Maximum and Minimum Display Values in Consideration of the Instrument’s Maximum Measuring and Intrinsic Uncertainties
- 25.2 At which values should/must an RCD actually be tripped? Requirements for Residual Current Devices (RCD)
- 25.3 Testing Electrical Machines per DIN EN 60 204 – Applications, Limit Values
- 25.4 Periodic Testing per DGUV V 3 (previously BGV A3) – Limit Values for Electrical Systems and Operating Equipment
- 25.5 Bibliography
- 25.6 Internet Addresses for Additional Information
Gossen Metrawatt GmbH 95
20 Test Sequences (Automatic Test Sequences)
– AUTO Function
Select AUTO Switch Position at the Test Instrument
With the rotary switch in the AUTO position, all of the test
sequences in the device are displayed.
If there aren’t any test sequences in the instrument, NO DATA
appears.
20.1 General (test sequence layouts)
If the same sequence of tests will be run frequently (one after the
other with subsequent report generation), for example as speci-
fied in the standards, it’s advisable to make use of test
sequences.
Automated test sequences can be compiled from manually cre-
ated individual measurements with the help of the test sequence
function.
A test sequence consists of up to 200 individual steps, which are
executed one after the other.
Fundamentally, differentiation is made amongst three types of
individual steps:
•Note (Visual Inspection test step) Test sequences are inter-
rupted when a pop-up message is displayed for the inspector.
The test sequences is not resumed until the message has
been acknowledged.
Sample Message Before Insulation Resistance Measurement
“Disconnect the device from the mains!”
• Visual inspection, testing and report generation: The test
sequence is interrupted when a passed/failed evaluation is
displayed. The comment and the results of the evaluation are
saved to the database.
• Measurement (“User-Evaluated Measurement” test step): same
as individual measurements with instruments with storage and
parameters configuration
20.2 Creation of Test Sequences with ETC
The test sequences are created at the PC with the help of ETC
software, and are then transferred to the test instrument.
Measurement parameters are also configured at the PC. How-
ever, parameters can be changed at the test instrument during
the test sequence before the respective measurement is started.
When the test step is started once more, the parameter settings
specified in ETC are loaded again.
ETC does not subject the parameters to a plausibility check. As a
result, the newly created test sequence should be checked at the
test instrument before it’s permanently added to the database.
Limit values are not currently set in ETC, and have to be adjusted
during the automatic test.
Accessing the Menu for Editing Test Sequences
In order to be able to edit existing test sequences (e.g. add test
steps or change parameter settings), they first have to be loaded
to ETC.
There are two ways to do this:
•ETC: Extras Test Sequences Load Test Sequences
(from the file “test_sequence_xyz.seq”)
or
• ETC: Device Test Sequences Receive Test Sequences
(from the connected test instrument)
Operating Overview: Creating Test Sequences at a PC
1 Create a new test sequence – enter a designation.
2 Change the designation of the selected test sequence.
3 Duplicate the selected test sequence,
(Copy) is added to the name of the duplicated sequence.
4 Delete the selected test sequence.
5 Create or add a new test step for the selected test sequence.
– Select the test step type from the list to this end and either accept or edit
its designation.
6 Duplicate the selected test step.
7 Delete the selected test step.
8 Change position of the selected test step within the sequence.
9 Select test parameters for the selected test step type from the list.
10 Select a setting for the measuring parameter from the list.
11 Accept change to the measuring parameter.
12 Exit the test sequences menu.
Saving Test Sequences in ETC to the PC
We recommend saving default test sequences, as well as edited
and new test sequences, to the PC or to other data
storage media using the desired filename (test_se-
quence_xyz.seq) with the help of the following menu command:
Extras Test Sequences Save Test Sequences. Data loss
resulting from certain administrative operations is prevented in this
way (see following notes).
Due to the fact that only up to 10 test sequences, can be trans-
ferred to the test instrument, no more than 10 test sequences
should be saved to any given file.
Test sequences which have been saved to a file can be reloaded
to ETC at any time by clicking “Extras Test Sequences Load
Test Sequences”.
Sequences can be further edited by clicking
“Extras Test Sequences Edit Test Sequences”.
Transferring Test Sequences from the PC to the Test Instrument
After executing the following ETC command, all previously cre-
ated test sequences (up to 10) are transferred to the connected
test instrument: “Device Test Sequences Send Test
Sequences”.
AUTO
1
2 3 4
5 6 7
8
109
11 12
Select the utilized test
instrument.
!