Manual

Table Of Contents
64 Gossen Metrawatt GmbH
15.6 Earthing Resistance Measurement, Mains Powered – Selective Earthing Resistance Measurement with Current Clamp Sensor
as Accessory
As an alternative to the conventional measuring method, measurement can also be performed with a current clamp sensor.
Key
R
B
Operational earth
R
E
Earthing resistance
R
L
Cable resistance
R
X
Earthing resistance through equipotential bonding
systems
R
S
Probe resistance
PAS Equipotential bonding busbar
RE Overall earthing resistance (R
E1
// R
E2
// water pipe)
Measurement without clamp: R
E
= R
E1
// R
E2
Measurement with clamp: R
E
= R
E2
=
Select Measuring Function
Select Operating Mode
Connection
To be connected: 2-pole adapter, clamp and probe
Set Parameters at Tester
Measuring range (test current selection):
1 k (40 mA), 100 (0.4 A), 10 (3.7 … 7 A)
In the case of systems with RCCBs, the DC offset and positive
half-wave (DC + ) functions can be selected (only in the
10 range and only with the METRAFLEX P300).
Connection: 2-pole adapter + clamp meter
After parameter selection: automatic setting to 10 measuring
range and 1 V/A or 1000 mV/A transformer ratio
Touch voltage: UL < 25 V, < 50 V, < 65 V, for information regard-
ing freely selectable voltage see section 10.8
Test current waveshape:
Sinusoidal (full-wave), DC offset and
positive half-wave (DC + )
System type: TN/TT, IT
Current clamp sensor transformation ratio: see table below
Set Parameters at Current Clamp Sensor
Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
Important Instructions for Use of the Current Clamp Sensor
Use only the METRAFLEX P300 or the Z3512A current clamp
sensor for this measurement.
Read and adhere to the operating instructions for
the METRAFLEX P300 current clamp sensor, as well as the
safety precautions included therein.
Observe direction of current flow (see arrow on the current
clamp sensor).
Use the clamp in the permanently connected state. The sen-
sor may not be moved during measurement.
The current clamp sensor may only be used at an adequate
distance from powerful extraneous fields.
Before use, always inspect the electronics housing, the con-
nector cable and the current sensor for damage.
In order to prevent electric shock, keep the surface of the
METRAFLEX clean and free of contamination.
Before use, make sure that the flexible current sensor, the con-
nector cable and the electronics housing are dry.
T
E
S
T
I
N
S
T
R
U
M
E
N
T
W
a
t
e
r
P
i
p
e
SE
2
E
1
B
U
Probe
I
Clamp
-----------------



R
E
Test Instrument METRAFLEX P300 Clamp Test Instrument
Parameters
Transformation Ratio
Switches Measuring
Range
Measuring Range
1:1
1 V / A
3 A (1 V/A) 3A 0.5 … 100 mA
01:10
100 mV / A
30 A (100 mV/A) 30 A 5 … 999 mA
1:100
10 mV / A
300 A (10 mV/A) 300 A 0.05 … 10 A