Measured Value Storage with Deviating Electrical Circuit Parameter
The electrical circuit parameter selected by yourself at the test instrument
does not coincide with the parameter entered under object data in the
structure.
Example: Residual operating current is specified as 10 mA in the data-
base, but you have performed measurement with 100 mA. If you want to
perform all future measurements with 100 mA, the value in the database
has to be changed by acknowledging with the key. The measured
value is documented and the new parameter is accepted.
If you want to leave the parameter in the database unchanged, press the
key. The measured value and the changed parameter are only docu-
mented in this case.
All
Please enter a designation (alphanumeric).
All
Operation with a Barcode Scanner
Error message when the “EDIT” entry field is opened and rechargeable
battery voltage is less than 8 V. Output voltage is generally switched off
during barcode scanner operation if U is less than 8 V in order to assure
that remaining battery capacity is adequate for entering designations for
devices under test and saving the measurement.
Remedy: Rechargeable batteries must be recharged or replaced towards
the end of their service life.
All
Operation with a Barcode Scanner
Current flowing through the RS 232 port is too high.
Remedy: The connected device is not suitable for this port.
All
Operation with a Barcode Scanner
Barcode not recognized, incorrect syntax.
All
Data cannot be entered at this location within the structure.
Remedy: Observe profile for preselected PC software (see SETUP menu in
section 8).
All
Measured value cannot be saved at this location within the structure.
Remedy: Make sure that you have selected the right profile for you PC
evaluation program in the SETUP menu (see section 8).
All
Memory is full.
Remedy: Save your measurement data to a PC and then clear memory at
the test instrument by deleting the database or by importing an empty
database.
All
Delete measurement or database elements.
This prompt window asks you to confirm deletion (YES).
SETUP
Attention!
!
Data loss after restoring default settings!
Back up your measurement data to a PC before pressing the respective
key.
This prompt window asks you to confirm deletion.