Error Messages — LCD Pictographs — PRO-AB Leakage Current Measuring Adapter
All
The parameters you have selected do not make sense in combination
with previously configured parameters. The selected parameter settings
will not be saved.
Remedy: Enter other parameters.
R
E
2-pole measurement via earthing contact plug is not possible in IT sys-
tems.
EXTRA
ta+I
The intelligent ramp is not possible with RCD types RCD-S and G/R.
Status
Test
Plug
Mea-
suring
Adapter
Function Switch
Position
Function/Meaning
AUTO
The test sequence includes a measurement which cannot be processed
by the connected test instrument. The corresponding test step must be
skipped. Example: The test sequence includes an RCM measurement
which has been sent to the PROFITEST MTECH+.
AUTO
The test sequence has been run successfully.
AUTO
No test sequences have been saved.
Cause: These may have been deleted as a result of any of the following
actions: changing the language, the profile or the DB mode,
or resetting the test instrument to its default settings.
Status
Test
Plug
Mea-
suring
Adapter
Function Switch
Position
Function/Meaning
EXTRA
I
L
Measuring range exceeded
Change to the larger measuring range (test instrument and leakage cur-
rent measuring adapter).
EXTRA
I
L
Test measurement:
The test has been passed.
The leakage current measuring adapter is now ready for use.
EXTRA
I
L
Test measurement:
The test has not been passed.
The leakage current measuring adapter is defective. Contact our repair
service department.
EXTRA
I
L
Test measurement:
Check the fuse in the leakage current measuring adapter.