Entry Plausibility Check – Parameters Combination Checking — LCD Pictographs
Status
Test
Plug
Mea-
suring
Adapter
Function Switch
Position
Function/Meaning
Parameter out of range
I
N
5 × 500 mA is not possible
I
N
/ I
F
EXTRA

ta + I∆
Types B, B+ and EV/MI not possible with G/R, SRCD, PRCD
>I
N
180° not possible for RCD-S, G/R, SRCD, PRCD-S, PRCD-K
I
N
/ I
F
DC not possible with G/R, SRCD, PRCD
I
N
/ I
F
Half-wave or DC not possible with type AC
I
N
/ I
F
EXTRA
RCM
DC not possible with type A, F
I
N
EXTRA
RCM
½ test current not possible with DC
I
N
2 × I
N
/ 5 × I
N
with full-wave only
R
E
DC+ with 10 only
R
E
No DC bias in the IT network
R
E
15 mA only possible in 1 k and 100 ranges!
EXTRA
RCM
With RCM: Types AC, F, B+ and EV/MI are not possible.
I
N
/ I
F
EXTRA
RCM
Measurement with half-wave or DC is not possible in IT systems.