Datasheet

Gossen Metrawatt GmbH 3
PROFITESTMASTER SERIES
PROFITESTMTECH+, MPRO, MXTRA, MBASE+
IEC 60364-6, EN 50110-1
Features of Instrument Variants Applicable Regulations and Standards
Nominal ranges of use
Voltage U
N
120 V (108 132 V)
230 V (196 253 V)
400 V (340 440 V)
Frequency f
N
16 ⅔ Hz
(15.4 18 Hz)
50 Hz (49.5 50.5 Hz)
60 Hz (59.4 60.6 Hz)
200 Hz (190 210 Hz)
400 Hz (380 420 Hz)
Overall voltage range
65 550 V
Overall frequency range 15.4 420 Hz
Line voltage Sinusoidal
Temperature range 0 C +40C
Battery voltage 8 12 V
Line impedance angle Corresponds to cos =1 0.95
Probe resistance < 50 k
PROFITEST …
(article number)
MBASE+
(M520S)
M
PRO
(M520N)
M
TECH+
(M522R)
M
XTRA
(M522P)
Testing of residual current devices (RCDs)
U
T
measurement without tripping the RCD
✓✓✓✓
Tripping time measurement
✓✓✓✓
Measurement of tripping current I
F
✓✓✓✓
Selective, SRCDs, PRCDs, type G/R
✓✓✓✓
AC/DC sensitive RCDs, types B and B+
——
✓✓
Testing of insulation monitoring devices (IMDs) ——
Testing of residual current monitoring devices (RCMs) ——
Testing for N-PE reversal
✓✓✓✓
Measurement of loop impedance Z
L-PE
/ Z
L-N
Fuse table for systems without RCDs
✓✓✓✓
Without tripping the RCD, fuse table ——
✓✓
15 mA measurement
1)
1)
The so-called live measurement is only advisable if there’s no bias current within
the system. Only suitable for motor protection switches with small nominal current
values.
✓✓✓✓
Earth resistance R
E
(mains operation)
I/U measuring method (2/3-wire measuring method via
measuring adapter: 2-pole/2-pole + probe)
✓✓✓✓
Earthing resistance R
E
(battery operation)
3 or 4-wire measuring method via PRO-RE adapter
Soil resistivity
E
(battery operation)
(4-wire measuring method
via PRO-RE adapter)
Selective earth resistance R
E (mains operation)
with 2-pole adapter, probe, earth electrode and current
clamp sensor
(3-wire measuring method)
✓✓✓✓
Selective earth resistance R
E (battery operation)
with probe, earth electrode and current clamp sensor
(4-wire measuring method
via PRO-RE adapter and current
clamp sensor
)
Earth loop resistance R
ELOOP
(battery operation)
with 2 clamps (current clamp sensor direct and current
clamp transformer via PRO-RE/2 adapter)
Measurement of equipotential bonding R
LO
Automatic polarity reversal
✓✓✓✓
Insulation resistance R
INS
Variable or rising test voltage (ramp)
✓✓✓✓
Voltage U
L-N
/ U
L-PE
/ U
N-PE
/ f
✓✓✓✓
Special measurements
I
L
, I
AMP
current measurement with clamp
✓✓✓✓
Phase sequence
✓✓✓✓
Earth leakage resistance R
E(INS)
✓✓✓✓
Voltage drop (U)
✓✓✓✓
Standing-surface insulation Z
ST
✓✓✓✓
Meter start-up (kWh test)
✓✓✓✓
Leakage current with PRO-AB (IL) adapter
——
Residual voltage test (Ures)
——
Intelligent ramp (ta + I)
——
Electric vehicles at charging stations (IEC 61851)
——
✓✓
Documentation of fault simulations at PRCDs with the
PROFITEST PRCD adapter
——
Features
Selectable user interface language
2)
2)
Currently available languages D, GB, I, F, E, P, NL, S, N, FIN, CZ, PL
✓✓✓✓
Memory (database for up to 50,000 objects)
✓✓✓✓
Automatic test sequence function
✓✓✓✓
RS-232 port for RFID/barcode reader
✓✓✓✓
USB port for data transmission
✓✓✓✓
ETC
PC database and report generating software
✓✓✓✓
Measuring category: CAT III 600 V / CAT IV 300 V
✓✓✓✓
DAkkS calibration certificate
✓✓✓✓
IEC 60364-6
EN 50110-1
Operation of electrical installations –
Part 100: General requirements
EN 60529 Test instruments and test procedures
Degrees of protection provided by enclosures (IP code)
IEC 60364-6 Low-voltage electrical installations –
Part 6: Tests
IEC 60364-7-710 Low-voltage electrical installations – Requirements for special
installations or locations –
Part 710: Medical locations
IEC 61010/
EN 61010/
Safety requirements for electrical equipment for measurement,
control and laboratory use
Part 1: General requirements (IEC 61010-1 + cor.)
Part 31: Safety requirements for hand-held probe assemblies for
electrical measurement and test (IEC 61010-031 + A1)
IEC 61140
DIN EN 61140
Protection against electric shock
Common aspects for installations and equipment
DIN EN 61326-1 Electrical equipment for measurement, control and laboratory
use – EMC requirements – Part 1: General requirements
IEC 61557/
EN 61557/
Electrical safety in low voltage distribution systems up to
1000 V AC and 1500 V DC – Equipment for testing, measuring
or monitoring of protective measures
Part 1: General requirements (IEC 61557-1)
Part 2: Insulation resistance (IEC 61557-2)
Part 3: Loop resistance (IEC 61557-3)
Part 4: Resistance of earth connection and equipotential
bonding (IEC 61557-4)
Part 5: Earthing resistance (IEC 61557-5)
Part 6: Effectiveness of residual current devices (RCDs) in TT,
TN and IT systems (IEC 61557-6) 6
Part 7: Phase sequence (IEC 61557-7)
Part 10: Electrical safety in low voltage distribution systems up
to 1000 V AC and 1500 V DC – Equipment for testing,
measuring or monitoring of protective measures
(IEC 61557-10)
Part 11: Effectiveness of type A and type B residual current
monitors (RCMs) in TT, TN and IT systems
(IEC 61557-11)
(PROFITEST MXTRA IQ only)
IEC 61851-1
DIN EN 61851-1
Electrical equipment for electric vehicles –
Electric vehicle conductive charging systems
– Part 1: General requirements