User Manual

Table Of Contents
GMC-I Messtechnik GmbH 47
Entry Plausibility Check – Parameters Combination Checking — LCD Pictographs
Err25
Parameter out of permissible range
Err26
I
N
5 x 500 mA is not possible
Err27
I
N
/ I
F
Types B/B+ and EV/MI not possible with G/R, SRCD, PRCD
Err28
I
N
180° not possible for G/R, SRCD, PRCD
Err29
I
N
/ I
F
DC not possible with G/R, SRCD, PRCD
Err30
I
N
/ I
F
Half-wave or DC not possible with type AC
Err31
I
N
/ I
F
DC not possible with type A, F
Err32
I
N
1/2 test current not possible with DC
Err33
I
N
2 x / 5 x IN with full-wave only
Err34
I
N
/ I
F
DC+ with 10 Ω only
Err35
R
E
15 mA only possible in 1 k and 100 ranges!
Err36
R
E
15 mA as loop measurement only
Err37
All
The parameters you have selected do not make sense in combination with previ-
ously configured parameters. The selected parameter settings will not be saved.
Remedy: enter other parameter settings.
Status
Error
No.
Position of the
Function Switch
Function / Meaning