Instructions
14 GMC-I Messtechnik GmbH
5 Characteristic Values
Measurement input PRO-SCHUKO plug
insert for PROFITEST
MTECH+ and MXTRA,
three 4 mm safety
sockets
PROFITEST PRIME
Mains connection Five 4 mm safety
sockets,
(adapter cable for
CEE 16A, CEE 32A
and CEE 63A avail-
able as accessory)
Ambient Conditions
Operating temperature
+5 ... + 45 C
Storage temperature -20 ... + 60 C
Relative humidity
Max. 75%, no conden-
sation allowed
Elevation Max. 2000 m
Power Supply
Auxiliary voltage 5 V via RS 232 port
Current consumption 60 mA (all relays
dropped out)
240 mA (2 relays
picked up)
600 mA (all relays
picked up)
Power consumption 3 W
Electrical Safety
Measuring category 300 V CAT III
Pollution degree 2
Protection class II per IEC 61010-1/
EN 61010-1/
VDE 0411-1
Fuse links F1 to F8:
8xF 8.0A H/500V
Test voltage Insulation test voltage
may not exceed
500 V.
Operating conditions Continuous operation
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2013,
class B
Interference immunity EN 61326-1:2013
EN 61326-2-1:2013
Mechanical Design
Protection Housing: IP 40 per
DIN VDE 0470, part
1 / EN 60 529
Table Excerpt Regarding Significance of
IP Codes
Dimensions
Housing (W x H x D):
approx. 27.5 x 10.5 x
12 cm
(with
PRO-SCHUKO
plug insert
)
Weight Approx. 1.1 kg
(test adapter without
interface cable)
Data Interface
Transmission
parameters 9600 baud,
1 stop bit, no parity
Transmission format All commands are
transmitted as ASCII
character strings.
IP XY
(1
st
digit X)
Protection
Against For-
eign Object In-
gress
IP XY
(2
nd
digit Y)
Protection Against
Water Ingress
4
1.0 mm dia.
0 Not protected
PROFITESTREMOTE Characteristic Values