7 Fault Simulation
7.1 Mode 2 Charging Cable
(single-phase)
Switch mains power on by
turning the top rotary selector switch (S1)
to the Status C position.
7.1.1 Simulated Interruption
Start with the bottom rotary selec-
tor switch (S2) in the first ON position in
the orange area for single-phase inter-
ruption.
Switch through the positions, one after
the other, in the clockwise direction (ta-
ble from top to bottom).
* The test object’s correct reaction to the simulat-
ed fault can be found in the manufacturer’s op-
erating instructions.
7.1.2 Simulated Wire Reversal
Turn the bottom rotary selector
switch (S2) to the L1-PE position in the
orange area for single-phase wire rever-
sal.
Instead of the L1 LED, the PE LED must light
up.
* The test object’s correct reaction to the simulat-
ed fault can be found in the manufacturer’s op-
erating instructions.
7.1.3 Simulation of PE to Phase –
PE-U
EXT
Set the bottom rotary selector switch
(S2) to the PE-U
EXT
position.
Interference voltage is switched to PE with
touch protection.
The L1 LED and the PE LED light up.
* The test object’s correct reaction to the simulat-
ed fault can be found in the manufacturer’s op-
erating instructions.
Step
Rotary
Switch
Tes t
Adapter
ICCB
Action
L1
LED
PE
LED
LED
1
*
2
3
4
5
6
Step
Rotary
Switch
Test
Adapter
ICCB
Action
L1
LED
PE
LED
LED
7
*
Step
Rotary
Switch
Test
Adapter
ICCB
Action
L1
LED
PE
LED
LED
8
*
PROFITESTEMOBILITY Fault Simulation