Datasheet

PROFITEST PVSUN and PVSUN MEMO
Test Instrument for Testing PV Modules
and Strings per DIN EN 62446 (VDE 0126-23)
2 GMC-I Messtechnik GmbH
Characteristic Values
Voltage Measurement U0
Measuring range 0 to 1000 V DC
(no transient voltages)
Resolution 1 V
Accuracy ±(2% rdg. + 1 d)
Current (direct)
Measuring range 0 to 20 A DC, measuring time < 1 s
Voltage range 2 to 1000 V DC
Resolution 0.1 A
Accuracy ± (1% rdg. + 1 d)
Overcurrent
protection max. 24 A (shutdown of internal circuit)
Insulation Resistance Measurement R
ISO
Earth Fault Measurement
Measuring range 0 to 1000 V DC
Resolution 1 V
Accuracy ± (5% rdg. + 1 d)
Low-resistance measurement
Measuring range 0 to 10 Ω
Test Current > 200 mA
Resolution 0.1 Ω
Accuracy ± (1% rdg. + 1 d)
Number of
measurements approx. 500 low-resistance measurements
(batteries: 1.5 V per IEC LR6)
Display
LCD Multiple display with background
illumination
Dot matrix: 128 x 64 pixels
Reference Conditions
Ambient temperature
+ 23 °C ± 2K
Relative humidity 40 to 75%
Battery voltage 6 V ± 1V
Ambient Conditions
Operating temperature
0 to 40 °C
Storage temperature –10 °C to 60 °C
Relative humidity < 80%, no condensation allowed
Elevation max. 2000 m above sea level
Power Supply
Batteries 4 ea. 1.5 V IEC LR6, AA, AM3, MN1500
Consumption approx. 20 μA when switched off
approx. < 30 mA during normal operation
approx. 190 mA with background
illumination
Electrical Safety
Voltage range 1000 V
Electromagnetic Compatibility (EMC)
EMC directive EMC 2004/108/EC
Basic standard EN 61326-1:2006
Data Memory (PROFITEST PVSUN MEMO only)
Memory capacity max. 10,240 data records
Memory type non-volatile flash memory
(data are retained even if
batteries are depleted)
Data Interface (PROFITEST PVSUN MEMO only)
Type USB 2.0 (USB 2.0 full speed (12 Mbps)
compatible)
Port mini USB type B (5-pin)
Cable USB 2.0 cable (A plug | mini B plug)
Mechanical Design
Protection Housing: IP 42
per DIN VDE 0470 part 1/EN 60529
Dimensions 209 x 98 x 35 mm
Weight approx. 500 g with batteries
Test voltage 250 V DC 500 V DC 1000 V DC
Measuring range 0.5 M
Ω to 1 MΩ 1 MΩ to 20 MΩ 1 MΩ to 20 MΩ
Resolution 0.1 MΩ 1 MΩ 1 MΩ
Accuracy ±(1% rdg. + 1 d) ±(1% rdg. + 2 d) ±(1% rdg. + 2 d)
Limit value > 0.5 M
Ω > 1 MΩ > 1 MΩ
Number of
measurements
approx. 1000 (with battery set per IEC LR6)