User manual
13
2. INTRODUCTION
The METRACLIP 41 / METRACLIP 410 current clamp
meters have been designed for reliable and accurate,
non-intrusive measurement of AC, DC and complex
waveform currents.
Using advanced Hall Effect technology the
METRACLIP 41 / METRACLIP 410 can measure
currents accurately from 5mA to 40 Amps
(METRACLIP 41) and a from 100mA to 400A
(METRACLIP 410).
Measurement features include:
True RMS reading
1mA / 10mA Resolution
Autoranging / Autozeroing / Auto Power Off
Data Hold
These features make it a powerful tool for use in
inverters, switch mode power supplies, industrial
controllers and other applications requiring current
measurement and/or waveform analysis.
3. SPECIFICATION
3.1 Electrical data
(All accuracies stated at 23°C ± 1°C)
METRACLIP 41
Measuring range ....................... 0 to 40A DC or AC
pk
Current Ranges (Autoranging) .. 4A / 40A
Resoution ................................. 1mA / 10mA
Overload capacity (60s) ............ 200A
METRACLIP 410
Measuring range ....................... 0 to 400A DC or AC
pk
Current Ranges (Autoranging) .. 40A / 400A
Resoution ................................. 10mA / 100mA
Overload capacity (60s) ……… 600A
3.2 General data
Overall DC accuracy ................. ±1% of rdg ±5 digits
Temperature Coefficient ........... ±0.1% of reading/°C
Frequency range (-0.5dB) ......... DC and 15 – 400Hz
Working voltage ........................ 300V AC
RMS
or DC
Operating temperature .............. 0°C to +50°C
Storage temperature with
battery removed ........................ -20°C to +85°C
Power supply ............................ 9 V Alkaline battery
PP3, MN 1604
or IEC6LR61