Data Sheet

3-349-392-03
3/4.14
GMC-I Messtechnik GmbH
METRACLIP62
Clip-On Ammeter for the Measurement of
Leakage and Charging Current
Applications
Measurement of leakage current at grounded conductors
Measurement of leakage current
at single-phase or 3-phase systems
Measurement of line current
Measurement of charging current
Applicable Regulations and Standards
Characteristic Values
Measuring sensor Hall effect
Measuring Ranges
1)
Frequency range: 50 60 Hz, corner frequency 2 kHz
2)
additional error for hall effect due to hysteresis: 3% rdg.
Key: rdg. = measured value (reading)
Reference Conditions
Ambient temperature
+23 C 5 C
relative humidity max. 80%
IEC 61010-1/EN 61010-1/
VDE 0411-1
Safety requirements for electrical equipment for
measurement, control and laboratory use
– General requirements
EN 60529
VDE 0470 Part 1
Test instruments and test methods
Degrees of protection provided by enclosures (IP
code)
DIN EN 61326
VDE 0843 Part 20
Electrical equipment for measurement, control
and
laboratory use – EMC requirements
Meas. Range Input Resolution
Accuracy at 18 ... 28
C,
max. 80% relative
humidity
A AC TRMS with automatic zero balancing
200 mA 0 199.9 mA 100 A 1.0% rdg. 5 digits
1)
2000 mA 0 1999 mA 1 mA 1.0% rdg. 5 digits
1)
10 A 0 9.99 A 0.01 A
1.0% rdg.
10 digits
1)
A DC with automatic zero balancing
200 mA 0 199.9 mA 100 A 1.0% rdg. 3 digits
2)
2000 mA 0 1999 mA 1 mA 1.0% rdg. 3 digits
2)
10 A 0 9.99 A 0.01 A 1.0% rdg. 10 digits
2)
3 measuring ranges 1 mA . . . 10 A AC TRMS / DC
High resolution due to hall effect sensor
Switch-over between DC and AC measurements
Zero balancing for DC measurements
Storage of reference values for AC measurements
Display of maximum or minimum current measured value
Storage of measured value (DATA HOLD)
Battery saving circuit
Insensitive to magnetic interference fields thanks to double
shielding