Manual
Table Of Contents
- 1 Safety Instructions
- 2 Applications
- 3 Documentation
- 4 Operating Overview
- 5 Initial Startup
- 6 System Settings
- 6.1 Digital Keyboard
- 6.2 Firmware
- 6.3 Language Selection
- 6.4 Time and Date Settings
- 6.5 Digital Display Settings
- 6.6 Set Display of Leading Zeros
- 6.7 Configuring Password Protection
- 6.8 Viewing the Instrument’s Name
- 6.9 Installing a Functions Expansion
- 6.10 Switching the Instrument Off
- 6.11 Default Settings (instrument reset)
- 7 Control Functions
- 8 Measurements
- 8.1 Enabling Parameter Changes
- 8.2 Insulation Resistance Measurement – RISO Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.3 Inter-Turn Short Circuit Measurement – COIL Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
- 8.4 Absorption Index Measurement – DAR (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.5 Polarization Index Measurement – PI (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6 Voltage Measurement
- 8.6.1 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter
- 8.6.2 Duty Cycle Measurement – Duty AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6.3 RPM Measurement – RPM AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6.4 Direct and Pulsating Voltage Measurement, V DC and V (AC+DC)
- 8.7 Resistance Measurement “W”
- 8.8 Capacitance Measurement F
- 8.9 Temperature Measurement with Resistance Thermometers – Temp RTD
- 8.10 Temperature Measurement with Thermocouple – Temp TC
- 8.11 Continuity Test
- 8.12 Diode Testing with Constant Current of 1 mA
- 8.13 Milliohm Measurement – Rlo (2-wire measurement) (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
- 8.14 Milliohm Measurement – mW/4 (4-wire measurement)
- 8.15 Current Measurement
- 8.15.1 Direct and Pulsating Current Measurement, Direct Connection, – A DC and A (AC+DC)
- 8.15.2 Alternating Current and Frequency Measurement, Direct Connection, – AAC and Hz
- 8.15.3 Direct and Pulsating Current Measurement with Current Clamp Sensor – ADC und A (AC+DC)
- 8.15.4 Alternating Current Measurement with Current Clamp Sensor – AAC and Hz
- 8.16 Measuring Sequences – Test Sequences
- 9 Interface Operation and Software
- 10 Characteristic Values
- 11 Maintenance and Calibration
- 12 Accessories
- 13 Returns and Environmentally Sound Disposal
- 14 CE Declaration
- 15 Addresses
Gossen Metrawatt GmbH 63
Response Time (after manual range selection)
Display
TFT color graphic display (55 x 36 mm) with analog and digital
display including unit of measure, type of current and various spe-
cial functions
Background Illumination
Activated background illumination can be regulated by means of
a light sensor.
Analog Bar Graph
Scaling Linear
Polarity display With automatic switching
Sampling rate 40 measurements per second and display
refresh
Digital Measured Value Display
Resolution /
character height 320 x 480 dots, 12 mm
Number of places 31000 / 3100
4¾-place in the
V, A, Hz and
measuring
functions depending on parameter setting
Overflow display “OL” is displayed for 31,000 digits
or 3100 digits
Polarity display “–” (minus sign) is displayed
if plus pole is connected to “”
Sampling rate 10 and 40 measurements per second
with
the Min-Max function except for the capaci-
tance, frequency and duty cycle measuring
functions
Refresh rate 2 times per second, every 500 ms
Electrical Safety
Protection category II per EN 6 010-1
Measuring category CAT III CAT IV
Nominal voltage 1000 V 600 V
Pollution degree 2
Test voltage 7.4 kV~ per EN 61 010-1
Fuses
Power Supply
Battery pack 3.7 V, 4000 mAh, LiPo
(approx. 25% self-discharging per year)
Operating time approx. 20 hours (without M
ISO
mea-
surement / R
Lo
/ R 4-wire measurement)
Battery indicator Battery capacity display via battery sym-
bol: , querying of
momentary charge level via menu function
Power OFF function
The multimeter is switched off automatically:
– when battery voltage
drops to below approx. 3.6 V
– if none of the keys or the rotary switch
are activated for an adjustable duration
of 10 to 59 min. and the multimeter
is not in the continuous operating mode
Rechargeable battery packs can only be recharged externally.
1
Times 0.7 for interface operation
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2013 class B
Interference immunity EN 61326-1:2013
Short-term measured value deviation of up to 10% may occur during electromag-
netic interference thus reducing the specified operating quality.
Ambient Conditions
Accuracy range 0 C ... +40 C
Operating temperature
(storage temp. with batt.)
10 C ... +50 C
20 C ... +50 C with rubber holster
Storage temperature 25 C ... +70 C (without battery)
Relative humidity 40 ... 75%, no condensation allowed
Elevation to 2000 m
Place of use
Indoors, except within specified ambient
conditions
Data Interface
Type
Bluetooth 4.2
Frequency band 2.402 ... 2.480 GHz,
Transmitting power Max. 91 mW
Functions – Query measuring functions
and parameters
– Query momentary measurement data
Measured Qty./
Measuring Range
Digital Display
Settling Time
Jump Function
of the Measured Quantity
V, V
A, A
1.5 s
From 0 to 80%
of upper range limit value
300 ... 3 M 2s
From to 50%
of upper range limit value
30 M,
M
@
U
ISO
Max. 5 s
Continuity < 50 ms
C (Pt 100) Max. 3 s
1.5 s
30 nF ... 300 FMax. 5s
From 0 to 50%
of upper range limit value
>10 Hz 1.5 s
Current measuring
ranges &
4-Wire m
measuring ranges
F1: FF 1 A/1000 V AC/DC,
6.3 x 32 mm
Fuse with breaking capacity of 30 kA at
1000 V AC/DC,
protects the current measurement
input in the 300 A to 1 A ranges
2-Wire m
measuring ranges
(METRAHIT IM XTRA BT and
METRAHIT IM E-DRIVE BT
only)
F2: FF 0.315 A / 1000 V
6.3 x 32 mm
Fuse with breaking capacity of 30 kA at
1000 V AC/DC
Measuring
Function
Nominal
Voltage U
N
DUT
Resistance
Operating
Time in
Hours
Number of Possible
Measurements with Nom.
Current per EN 61557
V 20
1
V 15
1
RISO
100 V 1 M 5
100 V 100 k 300
500 V 500 k 60
1000 V 2 M 20