8.15.2 Alternating Current and Frequency Measurement, Direct
Connection, – AAC and Hz
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
Set the rotary switch to A (A ).
Repeatedly press the Func softkey until the desired measuring
function is displayed.
If necessary, conduct zero balancing for AAC by pressing the
Zero softkey (see description below).
Safely connect the measuring instrument (without contact re-
sistance) in series to the power consumer as shown.
Switch supply power to the measuring circuit back on (3).
Read the display. Make a note of the measured value if the in-
strument is not being operated in the memory mode or the
transmission mode.
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any capaci-
tors.
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
Improving Accuracy by means of Zero Balancing – Zero
The momentarily measured current value can be subtracted from
future measurements in all measuring ranges.
Press the Zero softkey.
The value ascertained for Zero is saved and appears above the
measurement display. Automatic compensation is imple-
mented for future measurements.
The value for Zero is retained when the instrument is switched
to a different measuring function. The correction or offset value
is deleted by pressing Zero once again, or when the instrument
is switched off. The value is cleared from the display.