8.13 Milliohm Measurement – Rlo (2-wire measurement)
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
Disconnect supply power from the electrical circuit of the de-
vice to be measured and discharge all high-voltage capaci-
tors.
Attention!
!
The device under test must be voltage-free!
If interference voltage of Uext > 2 V is detected, visual
and acoustic warnings are generated. Measurement is
disabled as well.
Refer to section 8.6.4 regarding testing for the absence
of voltage with the help of the direct voltage measure-
ment.
Set the rotary switch to Rlo.
Select desired test current Ip set with the
 scroll keys.
Select desired test current Ip set polarity: /+/–
Connect the device under test as shown.
Activate the measurement by pressing the Start softkey on the
instrument or on the included probe.
If necessary, conduct zero balancing by pressing the Zero key
(see description below).
Press the Stop softkey in order to end the measurement.
Improving Accuracy by means of Zero Balancing – Zero
Cable resistance and contact resistance can be eliminated in all
measuring ranges by means of zero balancing.
Short circuit the measurement cables to this end.
Activate the measurement by pressing the Start softkey.
Insofar as the value for Zero is less than a permissible threshold
of 0 to 50% of the measuring range, the Zero softkey is no lon-
ger grayed out in the display and the Zero function can be acti-
vated by pressing the Zero key.
The value ascertained for Zero is saved and appears above the
measurement display. Automatic compensation of cable resis-
tance is activated for all subsequent measurements. Cable re-
sistance is deleted after the measurement has been ended.
The correction or offset value is deleted by pressing Zero once
again during measurement, or by ending the measurement.
The value is cleared from the display.
Note
The “Setup for present measurement” submenu cannot
be accessed for the milliohm measurement because zero
balancing and the polarity of test current I
P can be
selected directly in the measurement view.
Blown Fuse
In the event of a blown fuse, measurement is not possible and the
following display appears:
F2