8.3.1.1 Preparing for Measurement
Note
Inter-turn short circuit measurements may only be con-
ducted on voltage-free coils.
Set the rotary switch to “
RISO
” or “
COIL
”.
Repeatedly press the Func softkey until the measurement view
for COIL appears at the display.
Select the motor or coil type to be tested under the measure-
ment type parameter in “Setup for present measurement” (see
below).
Selecting the Type of Measurement
> Momentary Measurement Setup > Parameter 
Measurement Type
 Coil / 1-Ph. Motor / 3-Ph. Motor
3 x > Measurement View
1-Ph. Motor: AC motor (L1)
3-Ph. Motor: 3-Phase motor (U-V, U-W, V-W)
Coil Motor with up to 15 coils (L1 - L15)
Select polarity: unipolar or bipolar (see below).
Polarity Selection
> Momentary Measurement Setup > Parameter 
Polarity
 Bipolar / Unipolar
3 x > Measurement View
S– A
V
Tem p
m/4
COIL
S+
0V!
!
V
U
W
Example
COIL
COIL
TEST ADAPTER
with Star Connection
M
Connect
COIL TEST ADAPTER
and measurement cables
U
Surge max
1200 V!
RISO
COIL
V
U
W
L1
L3
L2
Example
with Delta Connection
Pulse
Current
up to 600 A
is Possible!