Gossen Metrawatt GmbH 23
Activating the Trigger
The trigger function cannot be set during memory mode opera-
tion.
Press the MENU key.
Press the General Setup softkey.
Select the Memory menu with the
 scroll keys.
Switch to the submenu with the help of the
scroll key.
Select the Trigger parameter with the  scroll keys.
Acknowledge the selected parameter by pressing the OK key.
Select the respective function (within, outside or off) with the
help of the
 scroll keys.
Acknowledge your change with the OK key. The entry cursor
once again marks the entire parameter line.
> General Setup
 Memory  Trigger
Adjusting the Trigger Threshold
These parameters cannot be set during memory mode operation.
Press the MENU key.
Press the General Setup softkey.
Select the Memory menu with the
 scroll keys.
Switch to the submenu with the help of the
scroll key.
Select the Trigger Low Limit or Trigger High Limit parameter with
the help of the  scroll keys.
Acknowledge the selected parameter by pressing the OK key.
Select the desired entry position within the parameter using
the
scroll keys and change the respective value with the
 scroll keys.
Acknowledge your change with the OK key. The entry cursor
once again marks the entire parameter line.
> General Setup
 Memory  Trigger Low Limit
+00000 digits / Trigger High Limit +00000 digits
Managing Groups
Groups can be created in order to be able to sort measured val-
ues when they’re saved. The groups can be managed.
Press the MENU key.
Press the General Setup softkey.
Select the Memory menu with the
 scroll keys.
Switch to the submenu with the help of the
scroll key.
Select the Groups parameter with the help of the
 scroll
keys.
Acknowledge the selected parameter by pressing the OK key.
Manage the groups with the help of the New, Edit and Delete
softkeys.
Entries are made via the digital keyboard (see section 6.1).
> General Setup
 Memory  Groups
Selecting a Group
Before starting the respective measurement, select a suitable
group from the list you’ve created to which the measured values
will be saved.
Press the MENU key.
Press the General Setup softkey.
Select the Memory menu with the
 scroll keys.
Switch to the submenu with the help of the scroll key.
Select the Groups parameter with the help of the
 scroll
keys.
Acknowledge the selected parameter by pressing the OK key.
Select the desired group with the
 scroll keys.
Acknowledge your selection with the Select softkey.
> General Setup
 Memory  Groups
 Select
7.7.3 Memory
Memory space in the instrument is limited to 300,000 measured
values. You can check memory occupancy and clear memory if
necessary.
Querying Memory Occupancy
You can query memory occupancy from within the “Info” menu
before, as well as during the save operation.
Memory occupancy range: 000.1% ... 099.9%.
> General Setup
 Info > Memory Occupancy x.x%
Clearing Memory (deleting measured values)
This function deletes all measured values from memory!
This function cannot be executed during memory mode opera-
tion.
> General Setup
 Memory  Clear Memory
The following security prompt appears before memory is cleared:
“Clear memory?”. This prompt must be acknowledged with “Yes”
via the function key (not with OK).
Confirmation is displayed at the end of the operation: “Memory
has been cleared.”