Manual
Table Of Contents
- 1 Safety Instructions
 - 2 Applications
 - 3 Documentation
 - 4 Operating Overview
 - 5 Initial Startup
 - 6 System Settings
- 6.1 Digital Keyboard
 - 6.2 Firmware
 - 6.3 Language Selection
 - 6.4 Time and Date Settings
 - 6.5 Digital Display Settings
 - 6.6 Set Display of Leading Zeros
 - 6.7 Configuring Password Protection
 - 6.8 Viewing the Instrument’s Name
 - 6.9 Installing a Functions Expansion
 - 6.10 Switching the Instrument Off
 - 6.11 Default Settings (instrument reset)
 
 - 7 Control Functions
 - 8 Measurements
- 8.1 Enabling Parameter Changes
 - 8.2 Insulation Resistance Measurement – RISO Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
 - 8.3 Inter-Turn Short Circuit Measurement – COIL Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
 - 8.4 Absorption Index Measurement – DAR (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
 - 8.5 Polarization Index Measurement – PI (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
 - 8.6 Voltage Measurement
- 8.6.1 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter
 - 8.6.2 Duty Cycle Measurement – Duty AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
 - 8.6.3 RPM Measurement – RPM AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
 - 8.6.4 Direct and Pulsating Voltage Measurement, V DC and V (AC+DC)
 
 - 8.7 Resistance Measurement “W”
 - 8.8 Capacitance Measurement F
 - 8.9 Temperature Measurement with Resistance Thermometers – Temp RTD
 - 8.10 Temperature Measurement with Thermocouple – Temp TC
 - 8.11 Continuity Test
 - 8.12 Diode Testing with Constant Current of 1 mA
 - 8.13 Milliohm Measurement – Rlo (2-wire measurement) (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
 - 8.14 Milliohm Measurement – mW/4 (4-wire measurement)
 - 8.15 Current Measurement
- 8.15.1 Direct and Pulsating Current Measurement, Direct Connection, – A DC and A (AC+DC)
 - 8.15.2 Alternating Current and Frequency Measurement, Direct Connection, – AAC and Hz
 - 8.15.3 Direct and Pulsating Current Measurement with Current Clamp Sensor – ADC und A (AC+DC)
 - 8.15.4 Alternating Current Measurement with Current Clamp Sensor – AAC and Hz
 
 - 8.16 Measuring Sequences – Test Sequences
 
 - 9 Interface Operation and Software
 - 10 Characteristic Values
 - 11 Maintenance and Calibration
 - 12 Accessories
 - 13 Returns and Environmentally Sound Disposal
 - 14 CE Declaration
 - 15 Addresses
 
Gossen Metrawatt GmbH 21
7.7 Measurement Data Recording
The multimeter offers the option of saving measurement data 
once by pressing a button or repeatedly as a measurement series 
over longer periods of time.
Data are stored to a battery-backed memory module, and are 
retained even after the multimeter is switched off. The system 
acquires measured values relative to real-time.
Note
Measurement data recording is ended automatically as 
soon as internal memory is full.
No measurement data are overwritten. Memory has to be 
cleared in order to record further measurement data (see 
page 23).
Stored measurement data can be read out via Bluetooth . See 
also interface operation in section 9 on page 56.
Refer to section 7.6 on page 20 concerning storage of current 
data to IZYTRONIQ software.
7.7.1 One-Time Only (manual) Storage
The STORE key is used for on-time only storage of a data value. 
Recording type must be set to “Data Value”.
➭ Press the MENU key.
➭ Press the General Setup softkey.
➭ Select the Memory menu with the 
 scroll keys.
➭ Switch to the submenu with the help of the 
 scroll key.
➭ Select the Recording Type parameter with the help of the 
 
scroll keys.
➭ Acknowledge the selected parameter by pressing the OK key.
➭ Select the Data Value parameter with the help of the 
 scroll 
keys.
➭ Acknowledge your change with the OK key. The entry cursor 
once again marks the entire parameter line.
 > General Setup 
 Memory   Recording Type 
Data Value 
 3x
7.7.2 Measurement Series (automatic storage)
Recording type must first be set to “Periodic” for measurement 
series, and several basic parameters have to be set, for example 
the sampling rate. 
Note
Consider available memory space when selecting your 
settings (see page 23).
Afterwards, you can start memory mode operation and then the 
measuring function at any time.
The general procedure is described here first, and the following 
subsections explain the respective parameters and their configu-
ration in detail.
➭ Activate recurrent recording:
➭ Press the MENU key.
➭ Press the General Setup softkey.
➭ Select the Memory menu with the  scroll keys.
➭ Switch to the submenu with the help of the 
 scroll key.
➭ Select the Recording Type parameter with the help of the 
 scroll keys.
➭ Acknowledge the selected parameter by pressing the OK 
key.
➭ Select the Periodic parameter with the help of the 
 
scroll keys.
➭ Acknowledge your change with the OK key. The entry cur-
sor once again marks the entire parameter line.
➭ Set the sampling rate for memory mode operation (see 
page 22).
➭ Set recording time (see page 22).
➭ Select an hysteresis in order to assure efficient use of memory 
capacity (see page 22).
➭ Select a trigger function if required (see page 22).
➭ Create a group if necessary in order to store the measured 
values in a sorted manner (see page 23).
➭ Check current memory occupancy (see page 23).
➭ Before starting lengthy measured value recordings, check the 
battery’s charge level (see section 5.1).
➭ Start recording.
➭ Press the MENU key.
➭ Press the General Setup softkey.
➭ Select the Memory menu with the 
 scroll keys.
➭ Switch to the submenu with the help of the  scroll key.
➭ Select the Recording Start parameter with the 
 scroll 
keys.
➭ Acknowledge the selected parameter by pressing the OK 
key. 
The “Start” setting changes to “Stop”. 
The following message appears: “Recording has started.” 
At the same time, “REC” appears in red to the left of the 
battery level display.
➭ The display is returned to the measuring function by 
pressing the ESC key three times.
➭ Select the desired measuring function and an appropriate 
measuring range.
➭ Perform the measurement.
➭ Stop recording:
➭ Press the MENU key.
➭ Press the General Setup softkey.
➭ Select the Memory menu with the 
 scroll keys.
➭ Switch to the submenu with the help of the  scroll key.
➭ Select the Recording Stop parameter with the 
 scroll 
keys.
➭ Acknowledge the selected parameter by pressing the OK 
key. 
The following message appears: “Recording has 
stopped”. The red “REC” symbol is cleared from the dis-
play.
➭ The display is returned to the measuring function by 
pressing the ESC key three times.
MENU
OK
ESC