Manual
Table Of Contents
- 1 Safety Instructions
- 2 Applications
- 3 Documentation
- 4 Operating Overview
- 5 Initial Startup
- 6 System Settings
- 6.1 Digital Keyboard
- 6.2 Firmware
- 6.3 Language Selection
- 6.4 Time and Date Settings
- 6.5 Digital Display Settings
- 6.6 Set Display of Leading Zeros
- 6.7 Configuring Password Protection
- 6.8 Viewing the Instrument’s Name
- 6.9 Installing a Functions Expansion
- 6.10 Switching the Instrument Off
- 6.11 Default Settings (instrument reset)
- 7 Control Functions
- 8 Measurements
- 8.1 Enabling Parameter Changes
- 8.2 Insulation Resistance Measurement – RISO Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.3 Inter-Turn Short Circuit Measurement – COIL Function (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
- 8.4 Absorption Index Measurement – DAR (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.5 Polarization Index Measurement – PI (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6 Voltage Measurement
- 8.6.1 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter
- 8.6.2 Duty Cycle Measurement – Duty AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6.3 RPM Measurement – RPM AC (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
- 8.6.4 Direct and Pulsating Voltage Measurement, V DC and V (AC+DC)
- 8.7 Resistance Measurement “W”
- 8.8 Capacitance Measurement F
- 8.9 Temperature Measurement with Resistance Thermometers – Temp RTD
- 8.10 Temperature Measurement with Thermocouple – Temp TC
- 8.11 Continuity Test
- 8.12 Diode Testing with Constant Current of 1 mA
- 8.13 Milliohm Measurement – Rlo (2-wire measurement) (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE only)
- 8.14 Milliohm Measurement – mW/4 (4-wire measurement)
- 8.15 Current Measurement
- 8.15.1 Direct and Pulsating Current Measurement, Direct Connection, – A DC and A (AC+DC)
- 8.15.2 Alternating Current and Frequency Measurement, Direct Connection, – AAC and Hz
- 8.15.3 Direct and Pulsating Current Measurement with Current Clamp Sensor – ADC und A (AC+DC)
- 8.15.4 Alternating Current Measurement with Current Clamp Sensor – AAC and Hz
- 8.16 Measuring Sequences – Test Sequences
- 9 Interface Operation and Software
- 10 Characteristic Values
- 11 Maintenance and Calibration
- 12 Accessories
- 13 Returns and Environmentally Sound Disposal
- 14 CE Declaration
- 15 Addresses
20 Gossen Metrawatt GmbH
7.6 Measured Value Memory – STORE Function.
The following options are available for the storage of measured
values:
• Store at the instrument by pressing the STORE key on the
instrument
• Store at the instrument by pressing the STORE key on the
probe (METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE
BT only)
• Store at the PC by triggering the PUSH/PRINT function in the
IZYTRONIQ report generating program
7.6.1 Remote read-out and storage via the Z270S probe
(METRAHIT IM XTRA BT and METRAHIT IM E-DRIVE BT only)
The probe with integrated control unit permits remote triggering at
difficult to access places, and at locations which require your full
attention. The probe can be used for all measuring functions
except for current measurement. The connector cable is shielded
against interference.
➭ Connect the probe’s double plug to the voltage socket (V).
➭ Connect the KS17-2 safety measurement cable to the ground
socket.
➭ Establish contact with the measuring point.
➭ Start the respective measuring function by pressing the START
key on the probe.
➭ As soon as the measured value has settled in, it can be stored
by pressing the STORE key on the probe.
Alternatively, the measured value can be stored with the help
of the STORE key on the instrument.
The REC storage symbol appears briefly in the header in order to
visualize the storage process.
Electrical Safety
Measurements per DIN EN 61010-031 may only be performed in
environments in accordance with measuring categories III and IV
with the safety cap attached to the control unit’s test tip.
In order to establish contact inside 4 mm sockets, the safety cap has
to be removed by prying open the snap fastener with a pointed
object (e.g. the other test probe).
7.6.2 Remote Read-Out and Storage via the PC – PUSH/PRINT
Function
The procedure for storage via the PUSH/PRINT function is
described in the online help included with IZYTRONIQ report gen-
erating software.
Maximum rated voltage 300 V
600 V
600 V
Measuring category
CAT IV CAT III CAT II
Maximum rated current 1 A 1 A 16 A
With safety cap attached —
Without safety cap — —
START
STORE
Test Tip
Safety Cap
Safety Collar
Z
2
7
0
S
REC
S– A
V
Tem p
m/4
COIL
S+
KS17-2