V/Hz, , Temperature, and A/Hz Measurements
5.7.4 Alternating Current Measurement with Current Clamp Sensors,
A AC and Hz
Transformer Output, Voltage/Current
When a current clamp sensor is connected to the multimeter
(V input (
METRAHIT E
BASE
: V input)), all current displays appear with
the correct value in accordance with the selected transformation fac-
tor. The only prerequisite is that the current sensor is equipped with
at least one of the below listed transformation factors, and that the
factor has been previously selected in the following menu (
Cl
ip

0FF
), see also chapter 6.4.
Current Clamp Setup Menu
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff

Maximum allowable operating voltage is equal to the current trans-
former’s nominal voltage. When reading the measured value, addi-
tional error resulting from the current clamp sensor must also be
taken into consideration.
(default setting:
METRAHIT E
XTRA
,
METRAHIT E
TECH
,
METRAHIT E
SPECIAL
:
OFF
,
METRAHIT E
BASE
:
1:1000
)
Trans. Factors
CL IP
Measuring Ranges DMM Clamp Types
600 mV 6 V
1:1
1mV/1mA
600.00 mA 6.0000 A
1:1
1mV/1mA
1:10
1mV/10mA
6.0000 A 60.000 A
1:10
1mV/10mA
1:100
1mV/100mA
60.000 A 600.00 A
1:100
1mV/100mA
1:1000
1 mV/1 A
600.00 A 6000.0 A
1:1000
1 mV/1 A
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
A~
Hz
003.50
A
AC
TRMS
A~
Hz
1:1000
050.10
Hz
AC
1:1000
FUNC
ENTER
COM
V
Tem p
V
~
R
x
Hz
Current
R
i
R
i
~ 1 M
METRAHIT EBASE
COM
V
Tem p
A
~
R
x
Current
R
i
R
i
= 1 M

9 M
mA
METRAHIT EBASE
V~
Hz
1M
1kHz
V~
R
i
= 1 M
R
i
= 9 M
METRAHIT EXTRA,
METRAHIT E
TECH,
METRAHIT E
SPECIAL
METRAHIT EXTRA,
METRAHIT E
TECH,
METRAHIT E
SPECIAL