Instructions
Table Of Contents
- 1 Safety Features and Precautions
- 2 Operating Overview – Connections, Keys, Rotary Switch, Symbols
- 3 Initial Start-Up
- 4 Control Functions
- 5 Measurements
- 5.1 Voltage Measurement
- 5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
- 5.1.2 Alternating Voltage Measurement with 1 MW Load Resistance and Frequency Measurement with Selectable Low-Pass Filter (METRA HIT Extra, METRA HIT Etech and METRA HIT Especial only)
- 5.1.3 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter (METRA HIT Extra, METRA HIT Etech and METRA HIT Especial only)
- 5.1.4 Frequency and Duty Cycle Measurements (METRA HIT Extra only)
- 5.2 Resistance Measurement, W
- 5.3 Continuity Test
- 5.4 Diode Testing with a Constant Current of 1 mA
- 5.5 Temperature Measurement
- 5.6 Capacitance Measurement (METRA HIT Extra and METRA HIT Etech only)
- 5.7 Current Measurement
- 5.7.1 Direct and Pulsating Current Measurement, Direct Connection, A DC and A (DC+AC) (METRA HIT Extra, METRA HIT Etech and METRA HIT Especial only)
- 5.7.2 Alternating Current and Frequency Measurement, Direct Connection, A AC and Hz (METRA HIT Extra, METRA HIT Etech and METRA HIT Especial only)
- 5.7.3 Direct and Pulsating Current Measurement with Current Clamp Sensors, A DC and A (DC+AC)
- 5.7.4 Alternating Current Measurement with Current Clamp Sensors, A AC and Hz
- 5.7.5 Direct, Pulsating and Current Measurement with Current Clamps Transformer A DC, A (DC+AC), A AC and Hz (METRA HIT Extra, METRA HIT Etech and METRA HIT Especial only)
- 5.1 Voltage Measurement
- 6 Device and Measuring Parameters
- 7 Interface Operation
- 8 Technical Data
- 9 Maintenance and Calibration
- 10 Accessories
- 11 Index
32 GMC-I Messtechnik GmbH
V/Hz, , Temperature, and A/Hz Measurements
Alternating Voltage Level Measurement (dB)
Voltage level measurement is used in order
to ascertain overall attenuation or boosting
of a transmission system (shown here as a
4-pole setup).
Voltage level [dB]
where U
1
= U
REF
(reference level)
Result > 1: boosting
Result < 1: attenuation
➭ Manually select the measuring range for the voltage amplitude.
When the instrument is switched to dB measurement, the pre-
viously selected voltage measuring range remains active.
➭ Repeatedly press the FUNC | ENTER multifunction key until unit
of measure dB appears at the display.
Lowest measurable frequencies and maximum permissible
voltages are included in chapter 8, “Technical Data”.
The level measurement function is now activated. The measured
value is calculated based upon the RMS value of the alternating
voltage component relative to the measuring range (600 mV
1000 V), and displayed.
The default setting for th reference level is 0 dB = 0,775 V (1 mW
to 600 ). This value can be adjusted in the „SET “ menu (see also
chapter 6.4):
1nFo
... SET rAtE ... dbrEF
0 .. 001 ... 99 V
Note
No terminal resistors have been integrated into the device. It
performs measurement with a high input impedance of
9M.
Input impedance for voltage measurement is listed under
technical data.
In order to be able to perform correct measurement at non-
terminated devices under test, the terminating resistor must
be connected to the terminals. Be sure to take power loss at
the terminating resistor into consideration!
U
1
U
2
20
U
2
U
1
------log=
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER