Current Measurement with Current Clamp Sensor
& Shunt Resistor
SR9800
Voltage/Current Transformer Output
When a current clamp sensor CP1800 is connected to the system
multimeter (mA/A input) via shunt resistor SR9800 (see Accesso-
ries, section 10.5) , all current displays appear with the correct
value in accordance with the selected transformation ratio. The
only prerequisite is that the current sensor is equipped with at least
one of the below listed transformation ratios, and that the ratio has
been previously selected in the following menu (
Cl
ip
0FF
) (see
also section 6.4).
Current Clamp Setup Menu
1nFo ... SET system ... CLIP
CLIP set: 1 :1E4
/
1E5
/
1E6
/
1E7 /0ff
The maximum permissible operating voltage is 300 V CAT III to
earth.
(default setting: Cl ip 0FF)
Trans. Ratio of
Clamp
Display /
SETUP Menu
CLiP
Current
Measuring
Ranges
Clamp Types
1:1 V/A
1:1 *10k
1:1E4
6,0000 A
1:10 V/A
1:10 *10k
1:1E5
60,000 A
1:100 V/A
1:100 *10k
1:1E6
600,00 A
CP1800
1:1000 V/A
1:1000 *10k
1:1E7
6000,0 A
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM
AmA
R
i
3
~
R
x
Strom
Current
R
i
SR9800
A
A
/
CLiP = 1:1/10/100/1000 * 10k (SETUP: 1:1E4/1E5/1E6/1E7)
60 mA / 600 mA / 6 A
bis 0,7 A dauernd
8
.
8
.
8
.
8
.
8
!
A measuring range:
Current measurement may
only be performed with
installed batteries!
10 A 5 min
CP1800