Device and Measuring Parameters
V AC with 30,000 digits), it is not useful to set the trigger thresh-
old above this measuring range span. It’s thus advisable to per-
form measurement with a fixed measuring range. Due to the fact
that rapid momentary value acquisition (see section 4.5.1) has a
large measuring range span, limit values of greater than 30,000
digits can be selected.
Actual measurement is always executed using the sampling rate
selected in “Store > rAtE”.
1nFo
... store ... trig
trig set : sto-ou / sto-in / off

l_trig set : -150000 ... +150000

h_trig set : -150000 ... +150000 
tStorE – Recording Time (parameter for memory mode operation)
This parameter determines whether or not measured values will
be recorded for a limited time only. If recording time will be limited,
its duration can be entered in hours, minutes and seconds. “On
means unlimited recording time.
1nFo
... store ... t .store
t .store set : on /99:00:00 (hh:mm:ss)

After completion of the recording time t.StorE, the end of the stor-
age process is signalled by 2 short acoustic signals (as from firm-
ware version V1.14).
6.6 Default Settings
Previously entered changes can be undone, and default settings
can be restored. This may be advisable under the following cir-
cumstances:
After the occurrence of software or hardware errors
If you are under the impression that the multimeter does not
work correctly
Disconnect the device from the measuring circuit.
Remove the batteries temporarily (see also section 9.2).
Simultaneously press and hold the and
keys and reinsert the batteries at
the same time, which is acknowledged with two acoustic
signals.