User manual

GMC-I Messtechnik GmbH 53
Device and Measuring Parameters
tEMP intErn/ExtErn – Select Internal or External Reference Junction
External Reference Junction: Specified Temperature
1nFo
... SET t iME ... tEMP intern
tEMP set: tEMP intern / tempextern

extern set: 000 . 0 °C

(internal reference = default setting)
bEEPEr on/oFF – Activate/Deactivate Acoustic Signals
1nFo
... SET t iME ... beeper off
beeper set: on / off

(bEEPEr on = default setting)
noFiL– quick display of measured values (as from firmwareversion 1.23)
For the following functions, a quicker display update rate of up to
5 display values per second (noFiL=ON) instead of 2 display val-
ues per second (noFiL=OFF) can be set:
V DC, A DC with clamp,
A DC direct,
, diode measurement.
1nFo
...
SET rAtE ... nofil
on / 0ff
(OFF =
Standard value*/default setting)
* All specifications refer to the default update rate with measuring param-
eter noFiL = OFF (default).
6.5 StorE Menu – Parameter for Memory Mode Operation
HYSt – Hysteresis (parameter for memory mode operation)
The hysteresis setting allows for efficient use of memory space.
During memory mode operation, new measured data are only
saved if they deviate from the previously stored value by an
amount which exceeds the selected hysteresis value.
Hysteresis can be selected in steps from 1 to 10,000 digits. These
digits are related to the measuring range as follows: The position
of the set digit in the specified hysteresis value corresponds to the
same position within the measuring range, although counting is
started at the left.
Example: A specified hysteresis of 00100 for the 600.00 V mea-
suring range means that only those measured values which devi-
ate from the last measured value by more than 001.00 V are
saved to memory.
Note
Due to the fact that the value is specified in digits (highest
place all the way to the left), and thus depends on the mea-
suring range, it’s advisable to use the function with a fixed
measuring range only.
1nFo ... store ... hyst
hyst set: 00000 ... 10000
