V/Hz, , Temperature, and A/Hz Measurements
5.7.1 Direct and Pulsating Current Measurement, Direct Connection, A
DC and A (DC+AC)
First disconnect supply power from the measuring circuit or
the power consumer, and discharge any capacitors.
In accordance with the current to be measured, set the rotary
switch to A or A
.
Select the type of current appropriate for the measured quan-
tity by briefly pressing the multifunction key FUNC | ENTER. Each
time the key is pressed, the instrument is switched back and
forth between A DC and A (DC + AC)
TRMS
, which is indicated
by means of an acoustic signal. The current type is indicated at
the LCD by means of the DC or the (DC+AC)
TRMS
symbol.
Safely connect the measuring instrument (without contact re-
sistance) in series to the power consumer as shown.
Switch supply power to the measuring circuit back on.
Read the display. Make a note of the measured value if the in-
strument is not being operated in the memory mode or the
transmission mode.
Disconnect supply power from the measuring circuit or the
power consumer once again, and discharge any capacitors.
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
Direct frequency measurement, A (DC + AC)
TRMS
(see follow-
ing section).
COM mA A
~
R
x
~
R
x
1
3
2
I > 10 A
!
A
A
300 A / 3 mA
30 mA / 300 mA
3 A / 10 A (10 A max. 5 min)
8
.
8
.
8
.
8
.
8
!
I > 16 A
Current
Measuring Ranges:
Current measurement
may only be performed
with installed batteries!
*
Additional temperature display
(internal temperature at jacks)
in the 3 A and 10 A ranges
(16 A max. 30 s)