User manual

64 GMC-I Messtechnik GmbH
Technical Data
Ambient Conditions
Accuracy range 0 C ... +40 C
Operating temp. range10 C ... +50 C
Storage temp. range 25 C ... +70 C (without batteries)
Relative humidity 40 to 75%, no condensation allowed
Elevation to 2000 m
Deployment
Indoors, except within specified ambient
conditions
Display
LCD panel (65 mm x 36 mm) with analog and digital display includ-
ing unit of measure, type of current and various special functions
Background illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Analog
Display LCD scale with bar graph or pointer,
depending upon the selected parameter
setting
Scaling Linear
(ranges other than M
ISO
):
5 ... 0 ... 30 with 35 scale divisions
for , 0 ... 30 with 30 scale divisions in all
other ranges
Logarithmic M
ISO
range):
 0.3 3 30 300
bar graph instead of pointer
Polarity display With automatic switching
Overflow display With the symbol
Measuring rate 40 measurements per second and display
refresh
Digital
Display / Char. Height7-segment characters / 15 mm
Number of places 3¾ places, 3100 steps
Overflow display “OL” is displayed for 3100 digits
Polarity display “–” (minus sign) is displayed
if plus pole is connected to “
Measuring rate 10 and 40 measurements per second with
the Min/Max function except for the
capacitance, frequency and duty cycle
measuring functions
Refresh Rate 2 times per sec., every 500 ms
Electrical Safety
Safety class II per DIN EN 61010-1:2011/
VDE 0411-1:2011
Measuring category II III
Nominal voltage 600 V 300 V
Fouling factor 2
Test voltage
3.5 kV~ per DIN EN 61010-1:2011/
VDE 0411-1:2011
Fuses
Fuse link FF 1.6 A/700 V AC/DC;
6.3 mm x 32 mm; switching capacity:
50 kA at 700 V AC/DC;
protects the current measurement input in
the 300 A through 1 A ranges