User manual

68 GMC-I Messtechnik GmbH
Technical Data
9 Technical Data
Calibrator Section
1)
The selection of frequencies as of 29 Hz is only possible within a limited raster.
Simulator for Temperature Sensors (resolution: 0.1 K)
2)
Without internal reference junction, relative to fixed external reference temperature
and thermovoltage of the thermocouple, internal reference junction: intrinsic error
of 2 K, external reference junction: entry of –30 60 C
Key
S = setting value
Calibration
Function
Simulation
Range
Resolution
30,000 Digits
(4¾ places)
Intrinsic
Uncertainty
Overload
Direct Voltage Source Minimum Load
Resistance
±(% S + mV) I
max
V
060 mV 1 μV
1k
0.1 + 0.01
18 mA
0300 mV 0.01 mV 0.05 + 0.02
0 3 V 0.1 mV 0.05 + 0.2
0 10 V 1 mV 0.05 + 2
0 15 V 1 mV 0.05 + 2
Pulse / Frequency Generator
Duty cycle (pulse-no-pulse ratio):
50%, Amplitude: 10 mV15 V
Minimum Load
Resistance
(% S + Hz) I
max
Hz 1 Hz 2 kHz 0,1 1 Hz 1 k 0.05 + 0.2 18 mA
Current Source Max. load (% S + A)
mA
4 20 mA
1 A 16 V 0.05 + 20 20 mA
0 24 mA
Current Sink (% S + A) U
max
mA
4 20 mA
1 AV
in
= 4 ... 26 V 0.05 + 2 26 V0 20 mA
0 24 mA
Resistance Simulation
Sensor Current [mA]
(% S + )I
max
52000 0.1 0.05...0.1..4...5 0.05 + 0.2 5 mA
Sensor Type Simulation
Range
in C
Simulation
Range
in F
Intrinsic
Uncertainty
Over-
load
C / F
Resistance Thermometer per IEC 751 (% S + K) I
max
Pt100 200 +850 328+1562 0.1 + 0.5
5 mA
Pt1000 –200 +300 –328 +572 0.1 + 0.2
Resistance Thermometer per DIN 43760 (% S + K) I
max
Ni100 60 +180 76 +356 0.1 + 0.5
5 mA
Ni1000 60 +180 –76 +356 0.1 + 0.2
RTD sensor current, 0.05 ... 0.1 ... 4
... 5 mA
Thermocouples per DIN and IEC 584-1 U in mV
2)
I
max
K (NiCr/Ni) –250+1372 418+2501
(0.05%
|Setting|
+
0.02)
18 mA
J (Fe/CuNi) 210+1200 346+2192
T (Cu/CuNi) –270+400 454+ 752
B (Pt30Rh/Pt6Rh) +500+1820 +932+3308
E (NiCr/CuNi) –270+1000 –454+1832
R (Pt13Rh/Pt) –50+1768 58+3214
N (NiCrSi-NiSi) –270+1300 –454+2372
S (Pt10Rh/Pt) –50+1768 58+3214
J (Fe/CuNi) 200+900 328+1652
U (Cu/CuNi) 200+600 –328+1112