V/Hz, , Temperature, and A/Hz Measurements
5.8.2 Alternating Current and Frequency Measurement,
Direct Connection, mA AC and Hz
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
Select the desired measured quantity by briefly pressing the
OUT | ENTER multifunction key. Each time the key is pressed,
AC
TRMS
and Hz are alternately selected, and switching is
acknowledged with an acoustic signal.
Safely connect the measuring instrument (without contact
resistance) in series to the power consumer as shown.
Switch supply power to the measuring circuit back on (3).
Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
~
R
x
~
R
x
1
3
2
mA~
mA~
Hz
003.50
A
AC
TRMS
050.10
Hz
AC
8
.
8
.
8
.
8
.
8
!
Current measurement only
with batteries installed!
Current
0.3 mA / 3 mA
30 mA / 300 mA
Measuring ranges:
ZERO/SEL
ESC
I > 310 mA
!
MR 300 mA
+ +
...