User manual

38 GMC-I Messtechnik GmbH
V/Hz, , Temperature, and A/Hz Measurements
5.8.1 Direct Current Measurement, A DC
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
Set the rotary switch to mA .
Safely connect the measuring instrument (without contact
resistance) in series to the power consumer (2) as shown.
Switch supply power to the measuring circuit back on (3).
Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
+ +
~
R
x
~
R
x
1
3
2
I > 310 mA
!
mA
003.50
A
DC
mA
0.3 mA / 3 mA
30 mA / 300 mA
8
.
8
.
8
.
8
.
8
!
Current
Measuring ranges:
Current measurement only
with batteries installed!
MR 300 mA
...