User manual

GMC-I Messtechnik GmbH 33
V/Hz, , Temperature, and A/Hz Measurements
5.6 Temperature Measurement
Temperature measurement is performed with a thermocouple
(accessory, not included), which is connected to the voltage in-
put. Alternatively, a resistance thermometer can be used.
Selecting the Unit of Measure for Temperature
1nFo
... setup time ...
temp unit °C / °F

(°C = default setting)
5.6.1 Measurement with Thermocouples, Temp TC
Set the rotary switch to “Temp
TC
”.
Note
The last selected temperature sensor remains in memory
and is accordingly displayed.
The reference temperature is measured at the internal
reference junction. It is shown in the right-hand auxiliary display
or else can be queried (see parameter “1tEMP” in section 7.3
regarding querying).
Note
The internal reference temperature (temperature of the
internal reference junction) is measured by a temperature
sensor inside of the instrument. This may be somewhat
above room temperature as a result of internal heat-up, or
moving from warmer to colder surroundings or vice versa.
Connect the sensor to the two accessible jacks. The
instrument displays the measured temperature using the
selected unit of measure.
MEAS / CAL
SETUP
+ +
Temp
TC
100.5
C
TC
TE (TC)
1
:
23
.
5
TC