6 Temperature Simulation [°C / °F]
The temperature simulator is capable of simulating resistance
temperature detectors (RTD) or thermocouples (TC) with
specification of the external reference junction temperature.
➭ Connect the device to be calibrated with the measurement
cables.
➭ Select the Temp RTD or Temp TC calibration function with the
rotary switch.
➭ Switch the calibrator on by pressing the ON / OFF | LIGHT
key.
The last selected temperature sensor is displayed.
➭ Set the temperature value:
Simulated resistance or simulated voltage is applied directly to the
output!
Select the digit to be changed with the Y Z keys, and change
the respectively selected digit with the VU keys. As an
alternative, press and hold the VU keys with the cursor at
any entry position until the higher value digits are scrolled
through as well.
➭ The output can be deactivated with the OUT | ENTER key
[
out.off], or activated once again.
Selecting Resistance Temperature Detector (RTD) or Thermocouple (TC)
for the Fixed Value, Interval or Ramp Function
➭ Switch to the fixed value, interval or ramp function menu by
pressing the SELECT | ESC key.
➭ Select the [
select sensor] menu with the YZ keys.
➭ Select the desired sensor with the VU keys. Acknowledge
your selection with OUT | ENTER. The display is switched to
the window for temperature value entry, but the selected
sensor still appears in the auxiliary display.
Query internal reference temperature – Info menu.
info
batt 2.9 V V ...V temp internal 23.7 °C
Parameter Entries for Thermocouple Temperature Simulation
Select unit of measure °C or °F – SEt menu.
info V set
time V ...V temp internal
unit set °F V °C
Select internal reference temperature – SEt menu.
info V set
time V ...V temp external
unit set temp set externa l V interna l
temp internal 23.7 °C
Select and set external reference temperature – SEt menu.
info V set
time V ...V temp internal
unit set temp set intern al V external
external set 21.0 °C YZ VU
temp external 22.4 °C
CAL
SETUP
OUT
ENTER
CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
CAL
SETUP
CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
CAL
SETUP
CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
CAL
SETUP