66 GMC-I Messtechnik GmbH
Accessories
10 Accessories
10.1 General
The extensive accessories available for our measuring instru-
ments are checked for compliance with currently valid safety reg-
ulations at regular intervals, and are expanded as required for new
applications. Currently up-to-date accessories which are suitable
for your measuring instrument are listed at the following web ad-
dress along with photo, order number, description and, depend-
ing upon the scope of the respective accessory, data sheet and
operating instructions: www.gossenmetrawatt.de
( Products Measuring Technology – Portable
Multimeters METRAHIT ... Accessories).
10.2 Technical Data for Measurement Cables
Electrical Safety of KS17-2 safety cable set
Please observe the maximum values of the electrical safety of the instrument.
Electrical Safety of KS17-2AMB safety cable set
(with antimicrobial properties, supply only with SECULIFEHITAM)
Please observe the maximum values of the electrical safety of the instrument.
Ambient Conditions (EN 61010-031)
Tem pera ture –20C ... + 50C
Relative humidity 50 to 80%
Pollution degree 2
Application KS17-2/KS17-2AMB
Attention!
!
Please observe the maximum values of the electrical supply
of the device.
In conformity with standard DIN EN 61010-031, measure-
ments in an environment according to measuring category
III and IV may only be performed with the safety cap applied
to the test probe of the measurement cable.
For establishing contact in 4 mm jacks you have to remove the
safety cap by levering out the snap lock of the safety cap with an-
other sharp object (e.g. the second test probe).
10.3 NA X-TRA Power Pack (not included)
Use only the power pack from GMC-I Messtechnik GmbH in
combination with your instrument. This assures operator safety by
means of an extremely well insulated cable, and safe electrical
isolation (nominal secondary ratings: 5 V / 600 mA). Installed
batteries are disconnected electronically if the power pack is
used, and need not be removed from the instrument.
Maximum Rated Voltage 600 V
1000 V
1000 V
Measuring Category
CAT IV CAT III CAT II
Maximum Rated Current 1 A 1 A 16 A
with safety cap applied
—
without safety cap applied ——
Maximum Rated Voltage 300 V
600 V
600 V
Measuring Category
CAT IV CAT III CAT II
Maximum Rated Current 1 A 1 A 16 A
with safety cap applied
—
without safety cap applied ——