Datasheet
GMC-I Messtechnik GmbH 5
METRA HIT AM BASE, AM PRO, AM TECH, AM XTRA
METRA HIT B
ASE, PRO, TECH, X-TRA
Advanced Multimeters
Power Supply
Battery 2 ea. 1.5 V mignon cell (2 ea. size AA),
alkaline manganese per IEC LR6
(2 ea. 1.2 V NiMH rechargeable battery
also possible)
Service life
with alkaline manganese: approx. 200 hours
Battery test Battery capacity display with battery
symbol in 4 segments:
.
Querying of momentary battery voltage via
me
nu function.
Power OFF function Multimeter is switched off automatically:
– If battery voltage drops to below prox. 1.8 V
– If none of the keys or the rotary switch
are activated for an adjustable duration
of 10 to 59 minutes, and the multimeter
is not in the continuous operation mode
Power pack socket
(METRA HIT AM XTRA / X-
TRA / OUTDOOR only)
If the NAX-
TRA power pack has been
plugged into the instrument, the batteries
are disconnected automatically.
Rechargeable batteries can only be
r
echarged externally.
Display
LCD panel (65 mm x 36 mm) with analog and digital display includ-
ing unit of measure, type of current and various special functions
Background illumination
Background illumination is switched off approximately 1 minute
aft
er it has been activated.
Analog
Display
LCD scale with bar graph or pointer, depend-
ing on the selected parameter setting
Scaling
With 4 division lines each, 1 bar/pointer cor-
responds to 500 digits at the digital display
Polarity display With automatic switching
Overflow display With the
symbol
Measuring rate 40 measurements per second and display
r
efresh
Digital
Display / char. height 7-segment characters / 15 mm
Number of places 4½ place
11,999 steps
Overflow display “OL” is displayed for 12
,000 digits
Polarity display “–” (minus sign) is displayed
if plus pole is connected to “”
Measuring rate 10 and 40 measurements per second
with
the Min-Max function except for the
capacitance, frequency and duty cycle
measuring functions
Refresh rate 2 times per sec., every 500 ms
Acoustic Signals
For voltage Intermittent signal at above 1000 V
For current Intermittent signal at above 10 A
continuous signal at above 16 A
Fuse
(except for METRA HIT AM BASE/METRA HIT B
ASE
)
Fuse FF (UR) 10 A/1000 V AC/DC;
10 mm x 38 mm,
Switching capacity: 30 kA at 1000 V AC/
DC, protects the current measurement
input in the 100 A through 10 A ranges
Electrical Safety
Per IEC 61 010-1:2010/VDE 0411-1:2011
Safety class
I I
Measuring category I I IIV
Operating voltage 1000 V 600 V
Pollution degree
2
Test v oltage
6.7 kV~
Electromagnetic Compatibility (EMC)
Interference emission EN 61 326-1: 2013, class B
Interference immunity EN 61 326-1: 2013
EN 61 326-2-1: 2013
Ambient Conditions
Accuracy range 0 C ... + 40 C
Operating temp. range T
A
10C ... + 50C
Storage temp. range 25C ...
+ 70C (without batteries)
Relative humidity 40 ... .75 %, no condensation allowed
only
METRA HIT O
UTDOOR
: max. 96 %
Elevation To 2000 m
Deployment
Indoors, except within specified ambient
conditions
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 200 x 87 x 45 mm
(without protective rubber cover)
Weight Approx. 0.35 kg with batteries
Protection Housing: IP 52 (pressure equalization by
means of th
e housing)
Extra for
METRA HIT O
UTDOOR
:
Housing: IP 65
Table excerpt regarding significance of the IP code
IP XY
(1
st
digit X)
Protection against pene-
tration of solid particles
IP XY
(2
nd
digit Y)
Protection against
penetr
ation by water
5 Dust protected 2 Dripping (15 in
clination)
6 Dust-proof 5 Jet-water