User Manual
Table Of Contents
- 1 Safety Features and Precautions
- 2 Operating Overview – Connections, Keys, Rotary Switch, Symbols
- 3 Initial Start-Up
- 4 Control Functions
- 5 Measurements
- 5.1 Voltage Measurement
- 5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
- 5.1.2 Alternating Voltage Measurement with 1 MW Load Resistance and Frequency Measurement with Selectable Low-Pass Filter (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.1.3 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.1.4 Frequency and Duty Cycle Measurements (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md only)
- 5.2 Resistance Measurement, W
- 5.3 Continuity Test
- 5.4 Diode Testing with a Constant Current of 1 mA
- 5.5 Temperature Measurement
- 5.6 Capacitance Measurement (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md and METRA HIT AM TECH only)
- 5.7 Current Measurement
- 5.7.1 Direct and Pulsating Current Measurement, Direct Connection, A DC and A (DC+AC) (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.7.2 Alternating Current and Frequency Measurement, Direct Connection, A AC and Hz (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.7.3 Direct and Pulsating Current Measurement with Current Sensor Clamp, A DC and A (DC+AC)
- 5.7.4 Alternating Current Measurement with Current Sensor Clamp A AC and Hz
- 5.7.5 Direct, Pulsating and Current Measurement with Current Transformer Clamp A DC, A (DC+AC), A AC and Hz (not METRA HIT AM BASE)
- 5.1 Voltage Measurement
- 6 Device and Measuring Parameters
- 7 Interface Operation (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md only)
- 8 Technical Data
- 9 Maintenance and Calibration
- 10 Accessories
- 11 Index
60 GMC-I Messtechnik GmbH
Technical Data
Electromagnetic Compatibility (EMC)
Interference emission
EN 61326-1:2013, class B
Interference immunity
EN 61326-1:2013
EN 61326-2-1:2013
Ambient Conditions
Operating
temp. range T
A
10 C ... +50 C
Storage
temp. range 25 C ... +70 C (without batteries)
Relative humidity Max.75%, no condensation allowed,
METRAHIT O
UTDOOR &
SECULIFEHIT
AM
/
MD
only:
max. 96 %
Elevation to 2000 m
Deployment
Indoors; outdoors only within specified
ambient conditions
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 200 x 87 x 45 mm
(without protective rubber cover)
Weight Approx. 0.35 kg with batteries
Protection Housing: IP 52
Extension of the
METRAHIT O
UTDOOR
&
SECULIFEHIT
AM
/
MD
:
Housing: IP 65
Table Excerpt Regarding Significance of the IP Code
IP XY
(1st digit X)
Protection against pene-
tration of solid particles
IP XY
(2nd digit Y)
Protection against
penetration by water
5 Dust protected 2 Dripping (15 inclination)
6 Dust-proof 5 Jet-water