User Manual
Table Of Contents
- 1 Safety Features and Precautions
- 2 Operating Overview – Connections, Keys, Rotary Switch, Symbols
- 3 Initial Start-Up
- 4 Control Functions
- 5 Measurements
- 5.1 Voltage Measurement
- 5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
- 5.1.2 Alternating Voltage Measurement with 1 MW Load Resistance and Frequency Measurement with Selectable Low-Pass Filter (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.1.3 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.1.4 Frequency and Duty Cycle Measurements (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md only)
- 5.2 Resistance Measurement, W
- 5.3 Continuity Test
- 5.4 Diode Testing with a Constant Current of 1 mA
- 5.5 Temperature Measurement
- 5.6 Capacitance Measurement (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md and METRA HIT AM TECH only)
- 5.7 Current Measurement
- 5.7.1 Direct and Pulsating Current Measurement, Direct Connection, A DC and A (DC+AC) (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.7.2 Alternating Current and Frequency Measurement, Direct Connection, A AC and Hz (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.7.3 Direct and Pulsating Current Measurement with Current Sensor Clamp, A DC and A (DC+AC)
- 5.7.4 Alternating Current Measurement with Current Sensor Clamp A AC and Hz
- 5.7.5 Direct, Pulsating and Current Measurement with Current Transformer Clamp A DC, A (DC+AC), A AC and Hz (not METRA HIT AM BASE)
- 5.1 Voltage Measurement
- 6 Device and Measuring Parameters
- 7 Interface Operation (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md only)
- 8 Technical Data
- 9 Maintenance and Calibration
- 10 Accessories
- 11 Index
50 GMC-I Messtechnik GmbH
Device and Measuring Parameters
dAtE – enter date
Entering the current date makes it possible to acquire measured
values in real-time.
1nFo
... SET rAtE ... dAtE
31.12 (DD: day . MM: month)
2005 (YYYY: year)
Date and time must be reentered after replacing the batteries.
tiME – set time
Entering the correct time makes it possible to acquire measured
values in real-time.
1nFo ... SET rAtE ... t iME
10:24 (hh:mm)
Date and time must be reentered after replacing the batteries.
CLIP – set transformation factor (clamp current factor)
See chapter 5.7.3 ff.
6.5 Default Settings
Previously entered changes can be undone, and the default
settings can be reactivated. This may be advisable under the
following circumstances:
• After the occurrence of software or hardware errors
• If you are under the impression that the multimeter does not
work correctly
➭ Disconnect the device from the measuring circuit.
➭ Remove the batteries temporarily (see also chapter 9.2).
➭ Simultaneously press and hold
the and keys,
and reinsert the batteries at the same time.
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
ZERO
ESC
ON / OFF
LIGHT