User Manual
Table Of Contents
- 1 Safety Features and Precautions
- 2 Operating Overview – Connections, Keys, Rotary Switch, Symbols
- 3 Initial Start-Up
- 4 Control Functions
- 5 Measurements
- 5.1 Voltage Measurement
- 5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
- 5.1.2 Alternating Voltage Measurement with 1 MW Load Resistance and Frequency Measurement with Selectable Low-Pass Filter (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.1.3 Alternating Voltage and Frequency Measurement V AC and Hz with Selectable Low-Pass Filter (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.1.4 Frequency and Duty Cycle Measurements (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md only)
- 5.2 Resistance Measurement, W
- 5.3 Continuity Test
- 5.4 Diode Testing with a Constant Current of 1 mA
- 5.5 Temperature Measurement
- 5.6 Capacitance Measurement (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md and METRA HIT AM TECH only)
- 5.7 Current Measurement
- 5.7.1 Direct and Pulsating Current Measurement, Direct Connection, A DC and A (DC+AC) (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.7.2 Alternating Current and Frequency Measurement, Direct Connection, A AC and Hz (METRA HIT AM X-TRA / AM TECH / AM PRO / Outdoor / SECULIFE HITam/md only)
- 5.7.3 Direct and Pulsating Current Measurement with Current Sensor Clamp, A DC and A (DC+AC)
- 5.7.4 Alternating Current Measurement with Current Sensor Clamp A AC and Hz
- 5.7.5 Direct, Pulsating and Current Measurement with Current Transformer Clamp A DC, A (DC+AC), A AC and Hz (not METRA HIT AM BASE)
- 5.1 Voltage Measurement
- 6 Device and Measuring Parameters
- 7 Interface Operation (METRA HIT AM XTRA / Outdoor / SECULIFE HITam/md only)
- 8 Technical Data
- 9 Maintenance and Calibration
- 10 Accessories
- 11 Index
GMC-I Messtechnik GmbH 43
V/Hz, , Temperature, and A/Hz Measurements
5.7.4 Alternating Current Measurement with Current Sensor Clamp
A AC and Hz
Transformer Output, Voltage/Current
When a current sensor clamp is connected to the multimeter V in-
put, METRAHITAM BASE: V input), all current displays appear with
the correct value in accordance with the selected transformation
factor.
The only prerequisite is that the current sensor is equipped with
at least one of the below listed transformation factors, and that the factor
has been previously selected in the following menu (
Cl
ip
0FF
), see also
chapter 6.4.
Current Clip Setup Menu
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff
Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the current sensor clamp must also
be taken into consideration.
(default setting: OFF,
METRAHITAM BASE: 1:1000)
Trans. Factors
CL IP
Measuring Ranges DMM Clip Types
100 mV 1 V 10 V
1:1
1 mV / 1 mA
100.00 mA 1.0000 A 10.000 A
WZ12C,
METRAFLEX
300M
1:10
1m V / 10 mA
1.0000 A 10.000 A 100.00 A
WZ12B, Z201A/B, ME-
TRAFLEX
1:100
1m V / 100 mA
10.000 A 100.00 A 1000.0 A
Z202A/B, Z203B,
METRAFLEX, Z204B
1:1000
1 mV / 1 A
100.00 A 1000.0 A (10000.0 A)
Z202A/B, Z203A/B,
WZ12C, Z204B,
METRAFLEX 3000
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
A~
Hz
003
.
50
A
AC
TRMS
A~
Hz
1:1000
050
.
10
Hz
AC
1:1000
FUNC
ENTER
COM
V
Tem p
V
~
R
x
Hz
Current
R
i
R
i
~ 1 M
COM
V
Tem p
A
~
R
x
Current
R
i
R
i
= 1 M
9 M
mA
METRAHIT AM BASE
V~
Hz
1M
1kHz
V~
R
i
= 1 M
R
i
= 9 M
METRAHIT AM PRO/TECH/XTRA/
O
UTDOOR/SECULIFE HIT
METRAHIT AM PRO/TECH/XTRA/
O
UTDOOR/SECULIFE HIT
METRAHIT AM BASE










