V/Hz, , Temperature, and A/Hz Measurements
5.7.3 Direct and Pulsating Current Measurement with Current Sensor
Clamp, A DC and A (DC+AC)
Transformer Output, Voltage/Current
When a current sensor clamp is connected to the multimeter V in-
put, METRAHITAM BASE: V input), all current displays appear with
the correct value in accordance with the selected transformation
factor.
The only prerequisite is that the current sensor is equipped with
at least one of the below listed transformation factors, and that the factor
has been previously selected in the following menu (
Cl
ip

0FF
), see also
chapter 6.4.
Current Clip Setup Menu
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff

Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the current sensor clamp must also
be taken into consideration.
(default setting: OFF,
METRAHITAM BASE: 1:1000)
Trans. Factors
CL IP
Measuring Ranges DMM Clip Types
100 mV 1 V 10 V
1:1
1 mV / 1 mA
100.00 mA 1.0000 A 10.000 A
1:10
1m V / 10 mA
1.0000 A 10.000 A 100.00 A
Z201A/B
1:100
1m V / 100 mA
10.000 A 100.00 A 1000.0 A
Z202A/B, Z203A/B,
Z204B
1:1000
1 mV / 1 A
100.00 A 1000.0 A (10000.0 A)
Z202A/B, Z203A/B,
Z204B
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM
V
Tem p
A
~
R
x
Current
003
.
50
A
1:1000
A
DC
003
.
50
A
1:1000
A
DC
AC
TRMS
A
A
FUNC
ENTER
R
i
R
i
= 9 M
COM
V
Tem p
V
~
R
x
Hz
Current
R
i
R
i
~ 1 M
METRAHIT AM BASE
mA
METRAHIT AM BASE
V
V
METRAHIT AM PRO/TECH/XTRA/
O
UTDOOR/SECULIFE HIT
METRAHIT AM PRO/TECH/XTRA/
O
UTDOOR/SECULIFE HIT