User manual

GMC-I Messtechnik GmbH 13
6.1 Low-Voltage Measurement
The instrument is equipped with a special 30 mV measuring range for measuring
voltage drop at fuses which is distinguished by high resolution (10 V) with a low
input resistance of 50 k.
Set the rotary switch to “Temp RTD”.
Select “V DC” measurement with probe by repeatedly pressing the FUNC
key until “mV DC” appears at the display.
Connect
the probe to the ” and “V” sockets.
7 Current Measurement
First disconnect supply power from the measuring circuit or the power
consumer and discharge all capacitors, if any are present.
Depending upon the type of current, set the rotary switch to “A ” or “A .
The symbol for the selected current type, (DC) or ~ (AC), appears at the LCD.
Connect the measuring instrument securely to the power consumer in se-
ries as shown in the diagram (without transition resistor).
Notes Concerning Current Measurement:
The instrument may only be
used in power installations if
the electrical circuit is pro-
tected with a fuse or a circuit
breaker with a rating of up to
20 A, and if nominal voltage at
the system does not exceed
240 V~ (AC) or 50 V (DC).
Set up the measuring circuit
in a mechanically secure
fashion such that it cannot
be inadvertently interrupted.
Use conductors with an
adequate cross section and
connectors of adequate
size in order to prevent
excessive warming.
In the A measuring ranges, an intermittent acoustic signal generates a warn-
ing if the measured value exceeds 10 A.
All measuring ranges up to 10 A are protected by a resetable, 15 A/
240 V AC/50 V DC auto-fuse. In order to conform to the CAT requirements,
an additional slow-blowing fuse link (T16A/500V) has been fitted in series
with the miniature circuit breaker which can only be replaced by service per-
sonnel once it is tripped.
If a fuse or circuit breaker in the active current measuring range is tripped,
FUSE appears at the digital display, and an acoustic signal is generated
simultaneously.