Datasheet

GMC-I Messtechnik GmbH 3
METRAport40S
Digital Multimeter
Display
LCD panel (95 x 40 mm) with analog and digital display including
unit of measure, type of current and various special functions
Type COG (chip on glass) for good legibility from
various directions
Background illumination
Background illumination (by means of LEDs) is activated with two
keys, and is switched off automatically after approximately 1
minute.
Analog
Display LCD scale with pointer
Scale length 80 mm for V and A ,
67 mm for all other ranges
Scaling 5 ... 0 ... 30 with 35 scale divisions
for ,
0 ... 30 with 30 scale divisions in all other
ranges
Polarity display With automatic switching
Overflow display With triangle
Measuring rate 20 measurements per second
Digital
Display / char. height 7-segment characters / 20 mm
Number of places 4¾ places 31000 steps
Overflow display “OL” appears
Polarity display “–” (minus sign) is displayed
if plus pole is connected to “
Measuring rate 2 measurements per second
Refresh rate
V (DC) , V~ (AC), A, , ,
C (Pt100, Pt1000) 2 per second
Hz 1 per second
C (K) 0.5 per second
Power Supply
Battery 2 ea. 1.5 V mignon cell,
alkaline manganese per IEC LR6,
zinc-carbon per IEC R6
Service life With alkaline manganese: approx. 200 h
With zinc-carbon: approx. 80 h
Battery test Battery capacity display with battery
symbol in 4 segments: “
Power saving circuit The device is switched off automatically:
– If the measured value remains
unchanged for a period of approximately
10 minutes, and if none of the controls
are activated during this time.
Automatic shutdown can be deactivated.
– If battery voltage drops to below
approx. 1.8 V
Fuses
Range
300 A to 10 A – Resetable auto-fuse
15 A, 240 V AC, 50 V DC
– A slow-blow fuse is additionally
connected in series to the auto-fuse, the
blowing or absence of which is detected
automatically (display “FUSE”):
T16A/500V AC, 6.3 mm x 32 mm
1.5 kA switching capacity at 500 V AC
and ohmic load
Electrical Safety
Safety class II per IEC 61010-1:2001/EN 61010-
1:2001/VDE 0411-1:2002
Measuring category CAT II
Operating voltage 300 V
Fouling factor 2
Test voltage 2.3 kV~ per IEC 61010-1:2001/
EN 61010-1:2001/VDE 0411-1:2002
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2006 class B
Interference immunity EN 61326-1:2006
EN 61326-2-1:2006
Ambient Conditions
Accuracy range 0 C ... +40 C
Operating temp. range10 C ... +50 C
Storage temp. range 25 C ... +70 C (without batteries)
Relative humidity Max. 75%, no condensation allowed
Elevation To 2000 m
Deployment Indoors, except
within specified ambient
conditions
USB Interface
The USB port is electrically isolated from the measuring circuit.
Operating voltage 5 V DC 10% from USB Port of PC
Current consumption
50 mA max, 25 mA typ.
USB-Interface
Type Mini-B, 5-pin, USB 1.1
Transfer 38400 Baud
parameters (1 Stopbit, no parity)
Pinning 1: VCC, 2: D–, 3: D+,
4: ID/not assigned, 5: GND