User manual
Table Of Contents
- 1 Safety Features and Precautions
- 2 Initial Start-Up
- 3 Selecting Measuring Functions and Measuring Ranges
- 4 LCD
- 5 Measured Value Memory “HOLD”
- 6 Saving Minimum or Maximum Values “MIN/MAX” Hold
- 7 Voltage Measurement
- 8 Current Measurement
- 9 Resistance Measurement
- 10 Continuity and Diode Testing
- 11 Temperature Measurement
- 12 Characteristic Values
- 13 Maintenance
- 14 Repair and Replacement Parts Service, Calibration Center* and Rental Instrument Service
- 15 Guarantee
- 16 Product Support
- 17 Recalibration Service

12 GMC-I Messtechnik GmbH
11 Temperature Measurement
The multimeter provides for the measurement of
temperatures within a range of – 50°C to + 400°C with the
help of a type K temperature sensor.
➭ Set the rotary selector switch to „°C“.
➭ Connect the sensor to the two accessible jacks.
The device indicates the measured temperature in C at
the digital display.
➭ Briefly press the FUNC key to switch between C and F.
Note!
The cold junction temperature (reference
temperature) is measured with a Pt100 temperature
sensor inside the instrument. It is displayed when the
measuring input is short-circuited. Due to internal
heating or a change from warm to cold environment
or vice versa, the reference temperature may differ
from the ambient temperature.
023.2
C
TC Measuring Range
Type_K –50.0 +400.0 C
C
FUNC
brief
TYPE K
C
F