User Manual

32 GMC-I Messtechnik GmbH
16.2 Fuses
The 10 A fuse interrupts the 10 A current range, and the 1.6 A
fuse protects the mA current measuring ranges. All other mea-
suring ranges continue to function.
When a fuse blows, eliminate the cause of the overload before
putting the instrument back into operation!
Replacing the Fuses
Unscrew the cover of the housing base from the top hous-
ing.
Remove the defective fuse from the fuse holders, e.g. with
the help of a probe and replace it with a new one.
Screw the cover of the base housing onto the top again.
Make sure that the new fuse makes good contact.
The following fuses are admissible:
for current measuring ranges up to 400 mA:
type FF 1.6 A/1000 V
~; 6.3 mm x 32 mm
for the 10 A current measuring range:
type FF 10 A/600 V
~; 6.3 mm x 32 mm
Attention!
!
It is imperative to ensure that only the fuses specified
above are installed!
If a fuse with different specifications is used, it may put
the operator, the system and the multimeter at risk.
The use of repaired fuses or the short-circuiting of the
fuse holder is not permissible.