User Manual
24 GMC-I Messtechnik GmbH
15 Technical Characteristics
Meas.
Func-
tion
Measuring
Range
Resolution Input
Impedance
Digital display
inherent devia-
tion at refer-
ence condition
+(...% rdg.
+...digits)
Overload Capacity
3)
V (AC) / V
(DC)
Overload Val-
ues
Over-
load
Dura-
tion
V
400.0 mV 100 μV > 20 MΩ 0.75 + 2
1050 V (DC)
Conti-
nuous
4.000 V 1 mV 11 MΩ
0.5 + 2
40.00 V 10 mV 10 MΩ
400.0 V 100 mV 10 MΩ
600 V 1 V 10 MΩ
V
400.0 mV 100 μV 11 MΩ 1.5 + 5
1050 V (AC)
rms
Conti-
nuous
4.000 V 1 mV 11 MΩ
1 + 540.00 V 10 mV 10 MΩ
400.0 V 100 mV 10 MΩ
600 V 1 V 10 MΩ 1 + 10
approx. voltage
drop at max.
meas. current
A
40.00 mA 10 μA 450 mV
0.8 + 2 480 mA
Conti-
nuous
400.0 mA 100 μA 4.2 V
10.00 A
1)
10 mA 750 mV 1.5 + 5 — —
A
40.00 mA 10 μA 450 mV
1 + 5 480 mA
Conti-
nuous
400.0 mA 100 μA 4.2 V
10.00 A
1)
10 mA 750 mV 2 + 5 — —
Open-circuit volt.
Ω
400.0 Ω 100 mΩ
approx. 0.45 V
0.8 + 5
500 V
DC/AC
rms
10
min
4.000 kΩ 1 Ω
0.8 + 240.00 kΩ 10 Ω
400.0 kΩ 100 Ω
4.000 MΩ 1 kΩ 1 + 5
40.00 MΩ 10 kΩ 2 + 5
400.0 Ω 100 mΩ Acoustic
signal for
0...< 75 Ω
(approx)
1.000 V 1 mV approx. 1 V 2 + 10