User Manual

20 GMC-I Messtechnik GmbH
12 Capacitance Measurement
Attention!
!
Absolutely verify that the device under test is electrically
dead. External voltages would falsify the measurement
results!
Set the function selector switch to „F“.
Connect the (discharged!) device under test to sockets „
and „F“ with measurement cables. Polarized capacitors must
be connected to the „ “ socket at the „–“ pole.
Note!
Resistors and semiconductor paths in parallel to the
capacitor falsify the measurement results!
To measure small value capacitors, please use short
measurement cables!
If the symbol is displayed, the measured values
are not valid.