4 Mains Connection
4.1 Connecting the Tester
āž­ Connect the tester to the 230 V mains with the mains plug (1). If no earthing contact
socket is available, or if only a three-phase socket is available, connection of the phase
conductors, the neutral conductor and the PE conductor can be accomplished with the
help of the adapter socket. It includes 3 permanently connected cables and is included
with the KS 13 accessory cable set.
Attention!
!
The instrument may only be connected to electrical supply systems with 230 V/
240 V which conform to the valid safety regulations (e.g. IEC 60364, VDE 0100) and
are protected with a fuse or circuit breaker with a maximum rating of 16 A.
The pick-off clips on the cables at the adapter socket may only be connected in the
voltage-free condition!
If mains voltage is present, characters are displayed at the LCD regardless of the position of
the measuring function selector switch, even if no DUT has been connected. Thus the pres-
ence of characters at the display indicates the presence of voltage – independent of measur-
ing function selector switch position.
The mains voltage value is indicated at the display in the ā€œU
Netz
250 Vā€ selector switch posi-
tion. In all other switch detent positions, characters are displayed which have no correlation
to a measured value, if a DUT has not been connected.
KS13
KS13
KS13
yellow-green
yellow-green
yellow-green
only possible in TN-C systems
L1
N
PE
L1
L2
L3
N
PE
L1
L3
N
L2