Maximum Allowable Protective Conductor Resistance Values Depending upon Cable Length
Under no circumstances may a value of 1 Ω be exceeded. The table is also valid for cable
reels and extension cables.
Attention!
!
The alligator clip (3) must make good contact with the housing of the device under
test!
The connector cable must be moved during measurement, section by section over
its entire length – for permanently installed devices only in so far as the cable is ac-
cessible during repair, modification or testing. If the resistance value fluctuates during
this manual test which is essential for continuity testing, it must be assumed that the
protective conductor is damaged, or that one of its connection points is defective.
Defects of this type must be corrected before any further tests are performed.
5.3.1 Special case: permanently connected instruments
A resistance value of up to 1 Ω measured between a suitable earthing contact and all
exposed conductive parts which might become charged with voltage in the event of an error
is allowable for permanently installed devices.
In the case of data processing systems or combinations of permanently installed individual
devices, the network should be decoupled and individual measurements should be per-
formed. If decoupling is not feasible, individual measurements may be performed at
interconnected devices.
Length to [m] 5 12. 5 20 27.5 35 42.5 50
more than 50
Max. resistance
[Ω]
0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
To housing for measurement of protective conductor resistance
R
SL