User manual

8 GMC-I Messtechnik GmbH
Reference Conditions
Positioning horizontal
Temperature +23 °C ± 2K
Relative Humidity 45 ... 55%
Measured Quantity
Frequency 45 ... 65 Hz
Measured Quantity
Waveshape sine
Deviation between
Effective and Rectified
Values < 0.5%
Battery Voltage 9 V ± 0.5 V
Power Supply
Batteries: 6 ea. 1.5 V single cell (6 x D-size) zinc-coal, type R20 or
alkaline-manganese, type LR 20, in accordance with IEC
Battery Service Life
Number of measurements possible with 6 ea. type R20 batteries (with
indicator lamp in test probe switched off): at least 3000 measurements of
an insulation resistance of 1 MΩ
(U
N
= 1000 V, 5 s measurement, 25 s pause and so forth)
Electrical Safety
Safety Class II
Nominal Voltage 1000 V
Test Voltage 5.55 kV~
Measuring category II
Pollution Degree 2
Electromagnetic Compatibility (EMC)
Interference Emission/
Interference Immunity EN 61326-1
Mechanical Design
Protection housing: IP 52
Dimensions 165 mm x 125 mm x 110 mm
Weight 1.6 kg with batteries
Making Capacity for the Measurement of Insulation Resistance
Electronic voltage limiting prevents voltage at the device under test from
substantially exceeding nominal voltage during the making operation (see
following characteristic curve).
8 Maintenance
Attention!
!
Completely disconnect the instrument from all external electrical
circuits before replacing the batteries or the fuse.
We recommend setting the function selector switch to the 1000 V
position during transport and maintenance, in order to prevent the
tester from being switched on inadvertently.
Ux
Rx [k
Ω
]
[V]
500
0
U
N
=500V
400
300
200
100
500
1000
R1 Range
Ux
Rx [M
Ω
]
[V]
500
0
U
N
=500V
400
300
200
100
10
20
R3 Range
600
Ux
Rx [M
Ω
]
[V]
500
0
U
N
=500V
400
300
200
100
0.5
1
R2 Range
600
1.5
2