Datasheet

GMC-I Messtechnik GmbH 3
METRISO 1000A
Insulation Tester
Reference conditions
Position horizontal
Ambient temperature +23 °C ± 2K
Relative humidity 45 ... 55 %
Frequency of quantity
under measurement 45 ... 65 Hz
(for voltage measurement)
Waveform of quantity
under measurement Sine
Deviation between effective
and rectified value < 0.5 %
Battery voltage 9 V ±0,5 V
Nominal conditions of use
Temperature 0 ... 40 °C
Positioning horizontal or vertical
Battery voltage 7 ... 10 V
Power supply
Batteries: 6 ea. 1.5 single cell (6 x D-Size)
Zinc-carbon type R20 or alcaline-manganese type LR 20 as per
IEC.
Service Life
Number of possible measurements with 1 set of type R20 batter-
ies (with lamp in test prod switched off): at least 3000 measure-
ments of an insulation resistance of 1 MΩ (U
N
= 1000 V, 5 sec.
measurement duration, switched off for 25 sec. etc.).
Electrical safety
Protection class II
Nominal voltage 1000 V
Tes t voltage 5.55 kV~
Measuring category II
Contamination level 2
Electromagnetic Compatibility (EMC)
Interference Emission EN 61326-1:2006 Class B
Interference Immunity EN 61326-1
Mechanical design
Protection type Housing: IP 52
Dimensions WxHxD Tester: 165 mm x 110 mm x 125 mm
Bag: 200 mm x 170 mm x 270 mm
Weight 1.6 kg with batteries
Included equipment
1 insulation tester,
1 carrying bag,
1 operating instructions