4.14.6 Trigger Function Selection
Example: Select „Tran“ in trigger function selection (Trig Function).
For the above operating procedures, the corresponding SCPI commands should be:
TRIG:FUNC TRAN Selects transient operation (TRAN) for trigger function
4.14.7 Knob Function
Example: Enable the knob function.
4.14.8 Key Sound
Example: Enable the key sounds.
4.14.9 Communication Interface
Example 1: Select RS232 interface; set the baud rate to 9600; set parity check to None; set data bit to 8; set stop bit to 2;
enable the flow control.
Example 2: Select GPIB interface, set the address to 18.
Procedure Description Display
1 Press the key to enter into the main menu.
Load Default :
Yes *No
2
Use the key to select the menu item “Trig Function”; use the knob or the key to select the parameter “Tran”.
Press the Enter key to confirm.
Trig Function :
*Tran List
3 Press the key to exit the main menu.
0.000V 0.000A
0.000
A CCH OFF
Procedure Description Display
1 Press the key to enter into the main menu.
Load Default :
Yes *No
2
Use the key to select the menu item “Knob”; use the knob or the key to select the parameter “On”.
Press the Enter key to confirm.
Knob :
On *Off
3 Press the key to exit the main menu.
0.000V 0.000A
0.000
A CCH OFF
Procedure Description Display
1 Press the key to enter into the main menu.
Load Default :
Yes *No
2
Use the key to select the menu item “Key Sound”; use the knob or the key to select the parameter “On”.
Press the Enter key to confirm.
Key Sound :
On *Off
3 Press the key to exit the main menu.
0.000V 0.000A
0.000
A CCH OFF
Steps Description Display
Step 1 Press the key to enter into the main menu.
Load Default :
Yes *No
Step 2
Use the key to select the menu item “Interface”; use the knob or the key to select the parameter “RS232”.
Press the Enter key to confirm.
Interface :
*RS232 USB GPIB
Step 3
Use the key to select the menu item “Baud Rate”; use the knob or the key to select the parameter “9600”.
press the Enter key to confirm.
Baud Rate :
*9600 19200
Step 4
Use the key to select the menu item “Knob”; use the knob or the key to select the parameter “On”.
Press the Enter key to confirm.
hlen Sie mit Hilfe der Taste den Menüpunkt „Parity Check“. Setzen Sie mit dem Drehgeber oder der Taste den Parameter
auf „None“. Betätigen Sie die Taste Enter, um die Eingabe zu bestätigen.
Parity Check :
*None Even Odd
Step 5
Use the key to select the menu item “Data Bit”; use the knob or the key to select the parameter “8”.
Press the Enter key to confirm.
Data Bit :
*8 7
Step 6
Use the key to select the menu item “Stop Bit”; use the knob or the key to select the parameter “2”.
Press the Enter key to confirm.
Stop Bit :
1 *2
Step 7
Use the key to select the menu item “Flow Control”; use the knob or the key to select the parameter “On”.
Press the Enter key to confirm.
Flow Control :
On *Off
Step 8 Press the key to exit the main menu.
0.000V 0.000A
0.000
A CCH OFF
Steps Description Display
Step 1 Press the key to enter into the main menu.
Load Default :
Yes *No
Step 2
Use the key to select the menu item “Interface”; use the knob or the key to select the parameter “GPIB”.
Press the Enter key to confirm.
Interface :
RS232 USB *GPIB
Step 3
Use the key to select the menu item “GPIB Address”; use Entry keys or the knob together with keys to select the
parameter “18”. Press the Enter key to confirm.
GPIB-Adresse :
18
Step 4 Press the key to exit the main menu.
0.000V 0.000A
0.000
A CCH OFF