4.14.4 Current Limit in CV Mode
Example: Set the load to CV mode; set the voltage value to 2 V; set the current limit value in CV mode (CV Curr Limit) to 20 A;
turn on the load. The detailed operations are shown below:
For the above operating procedures, the corresponding SCPI commands should be:
MODE CV Sets the mode
VOLT 2 Sets the voltage setpoint to 2 V
CV:CURR:LIM 20A Sets the current limit value to 20 A in CV mode
INPUT ON Turns on the load
4.14.5 Current Rise/Fall Rate in CC Mode
Example: Set the load to CCH mode, set the current value to 2 A; set the current rise rate (Curr Rise Rate) in CC mode to 0.002 A/μs;
set the current fall rate (Curr Fall Rate) in CC mode to 0.005 A/μs; turn on the load; turn off the load.
The detailed operations are shown below:
For the above operating procedures, the corresponding SCPI commands should be:
MODE CCH Sets the mode
CURR 2 Sets the current value to 2 A
CURR:RISE:RATE 0.002 Sets the current rise rate in CC mode to 0.002 A/μs
CURR:FALL:RATE 0.005 Sets the current fall rate in CC mode to 0.005 A/μs
INPUT ON Turns on the load
INPUT OFF Turns off the load
Procedure Description Display
1 Set the load to CV mode, and set the voltage value to 2 V (see section 4.5 for details).
0.000V 0.000A
2.000
V CV OFF
Step2 Press the key to enter into the main menu.
Load Default :
Yes *No
Step3
Use the ▼ key to select the menu item “CV Curr Limit”; use Entry keys or the knob together with ◄ ► keys to input the current
limit value to 20 A. Press the Enter key to confirm.
CV Curr Limit :
20.000A
Step4 Press the key to exit the main menu.
0.000V 0.000A
2.000
A CCH OFF
Step5 Press the key to turn on the load.
0.000V 0.000A
2.000
A CCH ON
Procedure Description Display
1 Set the load to CCH mode, and set the current value to 2 A (see section 5.1 for details).
0.000V 0.000A
2.000
A CCH OFF
2 Press the key to enter into the main menu.
Load Default :
Yes *No
3
Use the ▼ key to select the menu item “Curr Rise Rate”; use Entry keys or the knob together with ◄ ► keys to input the current
value to 0.002 A/μs. Press the Enter key to confirm.
Curr Rise Rate :
0.002
A/us
4
Use the ▼ key to select the menu item “Curr Fall Rate”; use Entry keys or the knob together with the ◄ ► keys to input the
current value to 0.005 A/μs. Press the Enter key to confirm.
Curr Fall Rate :
0.005
A/μs
5 Press the key to exit the main menu.
0.000V 0.000A
2.000
A CCH OFF
6 Press the key to turn on the load.
0.000V 0.000A
2.000
A CCH ON
7 Press the key to turn off the load.
0.000V 0.000A
2.000
A CCH OFF