4.14.2 Short Circuit Operation
Example: Short circuit operation in CV mode. The detailed operations are shown below:
The detailed operations are shown below:
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT ON Sets the load to short circuit
MODE CV Sets the mode
INPUT ON Turns on the load
4.14.3 Von Point/Von Latch
Example: Set the Von Latch to Off; set Von Point to 1 V; turn on the load; This example implements the automatic turning on/off of the
load‘s input, which simplifies the test operations greatly. The detailed operations are shown below:
For the above steps, the corresponding SCPI commands should be:
INP:VOLT:ON:LATCH OFF Disables the Von Latch
INP:VOLT:ON 1 Sets the Von voltage to 1 V
INPUT ON Turns on the load
Procedure Description Display
1 Press the key to enter into the main menu.
Load Default :
Yes *No
2
Use the ▼ key to select the menu item “Short”; use the knob or ◄ key to select the parameter “On”.
Press the Enter key to confirm.
Short
On *Off
3
Press the key to exit the main menu.
4 Press the key to enter into the mode selection and parameter setting menu; use ▲ ▼ keys to select CV mode.
Press the Enter key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
0.000V sCV OFF
5 Press the key to turn on the load.
0.000V 0.000A
0.000V sCV ON
Procedure Description Display
1 Press the key to enter into the main menu.
Load Default :
Yes *No
2
Use the ▼ key to select the menu item “Von Latch”; use the knob or ► key to select the parameter “Off”.
Press the Enter key to confirm.
Von Latch :
On *Off
3
Use the ▼ key to select the menu item “Von Point”; use Entry keys or the knob together with ▲ ▼ keys to input the voltage value.
Press the Enter key to confirm.
Von Point :
1.000
V
4 Press the key to exit the main menu.
0.000V 0.000A
0.000
A CCH OFF
5 Press the key to turn on the load.
0.000V 0.000A
0.000
A CCH ON